Automatic Common Mode Electrode-Amplifier Impedance Balance: Implementation and Results
- 1 September 2019
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 1-4
Abstract
Power-line interference (PLI) is the major disturbing factor in all ground-free biosignal acquisition applications. The PLI currents picked up by the body must flow through the sensing electrodes, then the electrode cables and finally via the amplifier input impedances they reach the amplifier ground. The electrode impedances and the amplifier input impedances form Wheatstone bridge. Because the electrode impedances vary with time, the bridge tends to be imbalanced and produces differential PLI which is amplified together with the useful signal. An automatic impedance balancing approach, using software PLL for line synchronization, was published. This paper validates the approach through practical results from its implementation.Keywords
This publication has 3 references indexed in Scilit:
- An Analog Bootstrapped Biosignal Read-Out Circuit With Common-Mode Impedance Two-Electrode CompensationIEEE Sensors Journal, 2018
- Automatic current driven electrode-amplifier impedance balance with SPLL synchronization2016
- Automatic common mode electrode-amplifier impedance balance with SPLL synchronization2016