On the correction, perturbation and modification of small field detectors in relative dosimetry
Authors:
- F TessierFTF Tessier
- J SeuntjensJSJ Seuntjens
- 11 September 2014
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 59 (19) , 5937-5952
Abstract
The purpose of this study was to derive a complete set of correction and perturbation factors for output factors (OF) and dose profiles. Modern small field detectors were investigated including a plastic scintillator (Exradin W1, SI), a liquid ionization chamber (microLion 31018, PTW), an unshielded diode (Exradin D1V, SI) and a synthetic diamond (microDiamond 60019, PTW). A Monte Carlo (MC) beam model was commissioned for use in small fields following two commissioning procedures: (1) using intermediate and moderately small fields (down to 2 × 2 cm(2)) and (2) using only small fields (0.5 × 0.5 cm(2) -2 × 2 cm(2)). In the latter case the detectors were explicitly modelled in the dose calculation. The commissioned model was used to derive the correction and perturbation factors with respect to a small point in water as suggested by the Alfonso formalism. In MC calculations the design of two detectors was modified in order to minimize or eliminate the corrections needed. The results of this study indicate that a commissioning process using large fields does not lead to an accurate estimation of the source size, even if a 2 × 2 cm(2) field is included. Furthermore, the detector should be explicitly modelled in the calculations. On the output factors, the scintillator W1 needed the smallest correction (+0.6%), followed by the microDiamond (+1.3%). Larger corrections were observed for the microLion (+2.4%) and diode D1V (-2.4%). On the profiles, significant corrections were observed out of the field on the gradient and tail regions. The scintillator needed the smallest corrections (-4%), followed by the microDiamond (-11%), diode D1V (+13%) and microLion (-15%). The major perturbations reported were due to volume averaging and high density materials that surround the active volumes. These effects presented opposite trends in both OF and profiles. By decreasing the radius of the microLion to 0.85 mm we could modify the volume averaging effect in order to achieve a discrepancy less than 1% for OF and 5% for profiles compared to water. Similar results were observed for the diode D1V if the radius was increased to 1 mm.Keywords
This publication has 21 references indexed in Scilit:
- Detector to detector corrections: A comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beamsMedical Physics, 2014
- Output correction factors for nine small field detectors in 6 MV radiation therapy photon beams: A PENELOPE Monte Carlo studyMedical Physics, 2014
- Monte Carlo-based diode design for correction-less small field dosimetryPhysics in Medicine & Biology, 2013
- Experimental analysis of general ion recombination in a liquid‐filled ionization chamber in high‐energy photon beamsMedical Physics, 2013
- Monte Carlo modelling of diode detectors for small field MV photon dosimetry: detector model simplification and the sensitivity of correction factors to source parameterizationPhysics in Medicine & Biology, 2012
- Monte Carlo linear accelerator simulation of megavoltage photon beams: Independent determination of initial beam parametersMedical Physics, 2011
- Implementing a newly proposed Monte Carlo based small field dosimetry formalism for a comprehensive set of diode detectorsMedical Physics, 2011
- Ionization chamber gradient effects in nonstandard beam configurationsMedical Physics, 2009
- The influence of small field sizes, penumbra, spot size and measurement depth on perturbation factors for microionization chambersPhysics in Medicine & Biology, 2009
- A new formalism for reference dosimetry of small and nonstandard fieldsMedical Physics, 2008