A Statistical Approach To Measure Uniformity Of Fuji Prescale Pressure Sensitive Film Samples
- 17 September 2021
- journal article
- research article
- Published by Bitlis Eren Universitesi Fen Bilimleri Dergisi in Bitlis Eren Üniversitesi Fen Bilimleri Dergisi
- Vol. 10 (3) , 920-929
Abstract
Fuji prescale pressure sensitive films (FPPSF) are generally used in biomechanical applications to measure the pressure distributions and the contact areas between two contact surfaces as in coloured stains. FPPSFs are usually calibrated based on the calibration methods using regression analysis to obtain a specific calibration curve. However, researchers in literature achieve different calibration curves based upon their tests and different processes. The purpose of this study is to examine the calibration stains (data) obtained experimentally using the FPPSF samples compressed between two smooth flat surfaces to indicate contact pressures. The gray-level co-occurrence matrix (GLCM) was used to evaluate the uniformity of the sample stains for quantitative evaluation of uniform distribution of the sample stains. Moreover, the generated calibration data is also compared and validated statistically with the related calibration curves presented in the available literature. As a result, the parameter of energy property of GLCM can be practicable to check the image uniformity of samples in order for performing the accuracy of a generated curve.Keywords
Funding Information
- Kocaeli Üniversitesi (2015/098)
This publication has 32 references indexed in Scilit:
- Characterization of the Performance of a Custom Program for Image Processing of Pressure Sensitive FilmJournal of Biomechanical Engineering, 2008
- Scaphoid Excision with Four-Corner Fusion: A Biomechanical StudyHAND, 2007
- Measuring contact area, force, and pressure for bioengineering applications: Using Fuji Film and TekScan systemsMedical Engineering & Physics, 2005
- Contact pressures in the flexed hip joint during lateral trochanteric loadingJournal of Orthopaedic Research, 2005
- Use of Pressure-Sensitive Film to Quantify Sources of Injury to FishNorth American Journal of Fisheries Management, 2005
- Evaluation of Nominal Contact Area and Contact Pressure Distribution in a Steel-Steel Interface by Means of Ultrasonic Techniques.JSME International Journal Series C, 2003
- An improved method for measuring tibiofemoral contact areas in total knee arthroplasty: a comparison of K-scan sensor and Fuji filmJournal of Biomechanics, 1999
- IMAGE-AVERAGING IN THE ANALYSIS OF DATA FROM PRESSURE-SENSITIVE FILMExperimental Techniques, 1997
- Sealing Fuji Prescale pressure sensitive film for protection against fluid damage: the effect on its responseStrain, 1995
- Contact areas and pressure distributions in the subtalar jointJournal of Biomechanics, 1995