Correction of matrix effects for As and Se in ICP OES using a Flow Blurring® multiple nebulizer
- 10 October 2012
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 27 (12) , 2132-2137
Abstract
A new and efficient multiple nebulizer based on Flow Blurring® hydrodynamics (FBMN) has been employed for elemental analysis. The aerosol generated by the FBMN has been characterized and compared with a conventional pneumatic nebulizer (MicroMist®, MM). Both nebulizers have been operated with the same cyclonic spray chamber for comparison. At a solution flow rate of 300 μL min−1 and a nebulizer gas flow rate of 0.60 L min−1 the aerosol produced by the FBMN is formed by droplets smaller than 53 μm, and the Sauter mean diameter (D3,2) and the median diameter (D50) of the primary aerosols are 16.3 and 16.8 μm, and 20.4 and 23.3 μm with the FBMN and MM, respectively. The sample introduction system based on the FBMN successfully corrects matrix effects in As and Se determination by inductively coupled plasma optical emission spectrometry. The FBMN makes feasible the implementation of an on-line internal standard (I.S.) strategy by mixing the I.S. solution with analytical reference solutions or a sample solution directly in the spray chamber. It has been demonstrated that the combination of FBMN and In as an internal standard leads to better accuracy for determining As in a lake sediment.Keywords
This publication has 14 references indexed in Scilit:
- Feasibility of the internal standardization in direct determination of arsenic in wine samples by inductively coupled plasma atomic emission spectrometryMicrochemical Journal, 2011
- Compensation for matrix effects on ICP-OES by on-line calibration methods using a new multi-nebulizer based on Flow Blurring® technologyJournal of Analytical Atomic Spectrometry, 2010
- Behaviour of a flow focusing pneumatic nebulizer with high total dissolved solids solution on radially- and axially-viewed inductively coupled plasma atomic emission spectrometryJournal of Analytical Atomic Spectrometry, 2006
- Flow focusing pneumatic nebulizer in comparison with several micronebulizers in inductively coupled plasma atomic emission spectrometryJournal of Analytical Atomic Spectrometry, 2006
- Enhanced liquid atomization: From flow-focusing to flow-blurringApplied Physics Letters, 2005
- Preliminary characterization and fundamental properties of aerosols generated by a flow focusing pneumatic nebulizerJournal of Analytical Atomic Spectrometry, 2004
- Investigation of non-spectroscopic interference and internal standardization method in axially and radially viewed inductively coupled plasma optical emission spectrometry using cross-flow and ultrasonic nebulizationJournal of Analytical Atomic Spectrometry, 2003
- Matrix-induced shift effects in axially viewed inductively coupled plasma atomic emission spectrometrySpectrochimica Acta Part B: Atomic Spectroscopy, 2001
- Sample Introduction Techniques for Atomic SpectroscopyAnalytical Chemistry, 1984
- Sample Introduction: The Achilles' Heel of Atomic Spectroscopy?Analytical Chemistry, 1984