Abstract
Objective:To develop an alpha dosimetry technique for activity calibration of alpha-emitting radiopharmaceuticals using the Gafchromic® EBT3 (Gaf-EBT3) radiochromic film (International Speciality product, Wayne, NJ).Methods:The Gaf-EBT3 has a tissue equivalent radiosensitive layer (approximately 28 μm) sandwiched between two 100-μm thick polyester sheaths, thereby making it insensitive to alpha particles. We have split a Gaf-EBT3 sheet using a surgical scalpel to remove one of the polyester protective layers and covered the radiosensitive layer with thin Mylar® foil (Goodfellow Cambridge Limited, Huntingdon, UK) (2.5 μm). Small pieces of modified film were exposed at contact with a 560-Bq thin 241Am source for 5, 10, 24 and 94 h. The optical density of the films was evaluated using an optical densitometer. The alpha energy spectra of the 241Am source were recorded using a Si(Li) surface barrier detector.Results:Time-integrated specific alpha surface activity (kBq cm−2 h) was represented as a function of o...

This publication has 15 references indexed in Scilit: