Rapid pacing and high‐frequency jet ventilation additively improve catheter stability during atrial fibrillation ablation
- 21 April 2020
- journal article
- research article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 31 (7) , 1678-1686
Abstract
Introduction: Catheter stability during atrial fibrillation ablation is associated with higher ablation success rates. Rapid cardiac pacing and high‐frequency jet ventilation (HFJV) independently improve catheter stability. Simultaneous modulation of cardiac and respiratory motion has not been previously studied. The objective of this study was to determine the effect of simultaneous heart rate and respiratory rate modulation on catheter stability.Methods: Forty patients undergoing paroxysmal atrial fibrillation ablation received ablation lesions at 15 prespecified locations (12 left atria, 3 right atria). Patients were randomly assigned to undergo rapid atrial pacing for either the first or the second half of each lesion. Within each group, half of the patients received HFJV and the other half standard ventilation. Contact force and ablation data for all lesions were compared among the study groups. Standard deviation of contact force was the primary endpoint defined to examine contact force variability.Results: Lesions with no pacing and standard ventilation had the greatest contact force standard deviation (5.86 ± 3.08 g), compared to lesions with pacing and standard ventilation (5.45 ± 3.28 g; P < .01) or to lesions with no pacing and HFJV (4.92 ± 3.00 g; P < .01). Lesions with both pacing and HFJV had the greatest reduction in contact force standard deviation (4.35 ± 2.81 g; P < .01), confirming an additive benefit of each maneuver. Pacing and HFJV together was also associated with a reduction in the proportion of lesions with excessive maximum contact force (P < .001).Discussion: Rapid pacing and HFJV additively improve catheter stability. Simultaneous pacing with HFJV further improves catheter stability over pacing or HFJV alone to optimize ablation lesions.Keywords
This publication has 24 references indexed in Scilit:
- Pacing Mediated Heart Rate Acceleration Improves Catheter Stability and Enhances Markers for Lesion Delivery in Human Atria During Atrial Fibrillation AblationJACC: Clinical Electrophysiology, 2018
- High frequency jet ventilation for motion management during ablation procedures, a narrative reviewActa Anaesthesiologica Scandinavica, 2017
- Contact Force in Atrial Fibrillation: Role of Atrial Rhythm and Ventricular ContractionsCirculation: Arrhythmia and Electrophysiology, 2015
- Relationship Between Catheter Contact Force and Radiofrequency Lesion Size and Incidence of Steam Pop in the Beating Canine HeartCirculation: Arrhythmia and Electrophysiology, 2014
- Paroxysmal AF Catheter Ablation With a Contact Force Sensing CatheterJACC, 2014
- Initial impedance decrease as an indicator of good catheter contact: Insights from radiofrequency ablation with force sensing cathetersHeart Rhythm, 2014
- Efforts to enhance catheter stability improve atrial fibrillation ablation outcomeHeart Rhythm, 2013
- Effect of respiration on catheter-tissue contact force during ablation of atrial arrhythmiasHeart Rhythm, 2012
- Importance of Catheter Contact Force During Irrigated Radiofrequency Ablation: Evaluation in a Porcine Ex Vivo Model Using a Force‐Sensing CatheterJournal of Cardiovascular Electrophysiology, 2010
- High-frequency jet ventilation: Utility in posterior left atrial catheter ablationHeart Rhythm, 2006