eGantryMate: A Piezo-Motor-Driven Lean and Flexible Assistance System for MR-Guided Interventions at 1.5T
Open Access
- 13 March 2024
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Access
- Vol. 12, 40267-40278
Abstract
In closed-bore MRI units, assistance systems play a crucial role in overcoming patient access limitations during percutaneous interventions. In this work, we present eGantryMate, a piezo-motor-driven assistance system specifically designed for MR-guided needle interventions in high-field MRI systems. eGantryMate consists of an instrument positioning unit and a control unit equipped with piezo motors, radiofrequency filters, and shielding. Paired with a real-time tracking sequence for automatic marker detection and projection of the instrument trajectory onto the MR image, eGantryMate enables precise and efficient needle interventions. Targeting experiments were performed by inserting a biopsy needle into a series of fiducial targets in a phantom, and usability experiments were conducted in vivo without needle insertion. The results show artifact-free MR imaging, minimal temperature rise on the instrument positioning unit, and precise targeting capabilities. These findings demonstrate eGantryMate’s ability to perform real-time needle alignments and insertions within the magnet bore, highlighting its potential to enhance the acceptability and efficacy of MR-guided interventions.Keywords
Funding Information
- German Federal Ministry for Economic Affairs and Energy (BMWi) through the Program “Zentrales Innovationsprogramm Mittelstand (ZIM)” (ZF4535603BA9)
- IraSME funding “E-GantryMate,”
- Open Access Publication Fund of the University of Freiburg
This publication has 45 references indexed in Scilit:
- Passive needle guide tracking with radial acquisition and phase‐only cross‐correlationMagnetic Resonance in Medicine, 2020
- Simultaneous orthogonal plane imagingMagnetic Resonance in Medicine, 2016
- Crushed rephased orthogonal slice selection (CROSS) for simultaneous acquisition of two orthogonal proton resonance frequency temperature mapsJournal of Magnetic Resonance Imaging, 2013
- Outer volume suppression in steady state sequences (OVSuSS) for percutaneous interventionsMagnetic Resonance in Medicine, 2011
- Automatic passive tracking of an endorectal prostate biopsy device using phase‐only cross‐correlationMagnetic Resonance in Medicine, 2008
- MR‐guided biopsy: A review of current techniques and applicationsJournal of Magnetic Resonance Imaging, 2008
- MR systems for MRI‐guided interventionsJournal of Magnetic Resonance Imaging, 2008
- MR‐guided intravascular interventions: Techniques and applicationsJournal of Magnetic Resonance Imaging, 2008
- Targeted‐HASTE imaging with automated device tracking for MR‐guided needle interventions in closed‐bore MR systemsMagnetic Resonance in Medicine, 2006
- Floating shield current suppression trapConcepts in Magnetic Resonance Part B, 2004