Effect of detector lag on CT noise power spectra
Open Access
- 27 May 2011
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 38 (6Part1) , 2995-3005
Abstract
The authors examined the effect of detector lag on the noise power spectrum (NPS) of CT images reconstructed with filtered backprojection (FBP). The authors derived an analytical expression of the NPS with detector lag, and then verified it using computer simulations with parallel beam and fan beam geometries. The dependence of the NPS on the amount of lag, location within the scanned field of view (FOV), and the number of views used in the reconstruction (samples per rotation) was investigated using constant and view dependent noise in the raw data. Detector lag introduces noise correlation in the azimuthal direction. The effect on the NPS is a frequency dependent reduction in amplitude. In small regions of the image, the effect is primarily in the frequencies corresponding to the azimuthal direction. The noise blurring and NPS filtering increases with increasing radial distance, and therefore regions at larger radial distances have lower noise power. With the same detector lag response function, the amount of noise correlation and NPS filtering decreases with increasing number of views. The shape of the NPS depends on the detector lag coefficients, location of the region, and the number of views used in the reconstruction. In general, the noise correlation caused by detector lag decreased the amplitude of the NPS.Keywords
Funding Information
- National Institutes of Health (EB006837)
This publication has 28 references indexed in Scilit:
- Local and global 3D noise power spectrum in cone‐beam CT system with FDK reconstructionMedical Physics, 2011
- The noise power spectrum in CT with direct fan beam reconstructionMedical Physics, 2010
- A small-signal approach to temporal modulation transfer functions with exposure-rate dependence and its application to fluoroscopic detective quantum efficiencyMedical Physics, 2009
- An empirical method for lag correction in cone‐beam CTMedical Physics, 2008
- Signal behavior of polycrystalline HgI2at diagnostic energies of prototype, direct detection, active matrix, flat-panel imagersPhysics in Medicine & Biology, 2008
- Computer modeling of the spatial resolution properties of a dedicated breast CT systemMedical Physics, 2007
- Systematic investigation of the signal properties of polycrystalline HgI2detectors under mammographic, radiographic, fluoroscopic and radiotherapy irradiation conditionsPhysics in Medicine & Biology, 2005
- A framework for noise‐power spectrum analysis of multidimensional imagesMedical Physics, 2002
- Investigation of a solid-state detector for advanced computed tomographyIEEE Transactions on Medical Imaging, 2000
- The noise power spectrum in computed X-ray tomographyPhysics in Medicine & Biology, 1978