An Ultra-Low Power Fixed-Window Level Crossing ADC for ECG Recording
- 12 March 2024
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Circuits and Systems
- Vol. 18 (5) , 1089-1099
Abstract
In this paper, a novel fixed-window level-crossing analog-to-digital converter (LCADC) is proposed for the ECG monitoring application. The proposed circuit is implemented using fewer comparators and reference levels compared to the conventional structure, which results in a decrease in complexity and occupied silicon area. Also, the power consumption is reduced considerably by decreasing the activity of the comparator. Simulation results show a 5-fold reduction in activity by applying the standard ECG signals to the proposed structure. The proposed circuit is implemented in 0.18 μm CMOS technology using a 0.9 V supply voltage. Measurement results show a 5.9 nW power consumption and a 7.4-bit resolution. The circuit occupies a 0.05846 mm 2 silicon area. A typical level-crossing-based R-peak-detection algorithm is applied to the output samples of the LCADC, which shows the effectiveness of using this type of sampling.Keywords
This publication has 33 references indexed in Scilit:
- Reducing the power consumption of level-crossing analog-to-digital converters by using a signal activity monitoring circuitMicroelectronics Journal, 2023
- A 200 Hz-to-10 kHz bandwidth 11.83-ENOB level-crossing ADC with single continuous-time comparatorMicroelectronics Journal, 2022
- An Ultra-Low Power Level-Crossing ADC for ECG Monitoring Application2020
- Level-crossing sampling: principles, circuits, and processing for healthcare applications2020
- A 10-kS/s 625-Hz-Bandwidth 65-dB SNDR Second-Order Noise-Shaping SAR ADC for Biomedical Sensor ApplicationsIEEE Sensors Journal, 2019
- Reconstruction of Signals From Level-Crossing Samples Using Implicit InformationIEEE Access, 2018
- A 2.2- $\mu$ W Cognitive Always-On Wake-Up Circuit for Event-Driven Duty-Cycling of IoT Sensor NodesIEEE Journal on Emerging and Selected Topics in Circuits and Systems, 2018
- Ultra-low-power fully-synchronous level-crossing analog-to-digital converter for biomedical signal acquisition2017
- A Level-Crossing Based QRS-Detection Algorithm for Wearable ECG SensorsIEEE Journal of Biomedical and Health Informatics, 2014
- A Continuous-Time ADC/DSP/DAC System With No Clock and With Activity-Dependent Power DissipationIEEE Journal of Solid-State Circuits, 2008