UV light-based reprocessing of flexible endoscopes without working channel in Oto-Rhino-Laryngology: an effective method?
Open Access
- 13 March 2021
- journal article
- research article
- Published by Springer Nature in European Archives of Oto-Rhino-Laryngology
- Vol. 278 (10) , 4075-4080
Abstract
Background: Reprocessing of flexible endoscopes (FEs) is often expensive, time consuming, and becomes increasingly complex, due to rising demands of hygiene. After beneficial results in reprocessing of rigid endoscopes using Impelux™ UV-C light technology, we tested the same method for reprocessing of FEs without working channel. Materials and methods: Testing was performed on FEs without working channel after routine clinical use (transnasal flexible endoscopy). Disinfection consisted of mechanical precleaning and 60 s exposure to Impelux™ UV-C light technology. Bacterial contamination was tested on 50 FEs before and after disinfection. Further 50 FEs regarding protein residuals. The absolute effectiveness of the D60 was tested on 50 test bodies (RAMS) with a standardized contamination of 107 colony-forming units (CFU) of Enterococcus faecium. Results: The FEs were contaminated with a high average value of 916.7 CFU (± 1057 CFU) after clinical usage. After reprocessing, an average contamination of 2.8 CFU (± 1.6) on 14% (n = 7) of the FEs was detected consisting of non-pathogenic species, the remaining FE were sterile. After reprocessing, all FEs were protein-free (< 1 μg). The artificially contaminated test bodies showed no remaining bacterial contamination after disinfection, resulting in an average absolute germ reduction of about 107 CFU. Conclusion: Impelux™ UV-C light technology efficiently reduces bacterial contamination of FEs and might be useful in daily practice.Keywords
Funding Information
- Philipps-Universität Marburg
This publication has 20 references indexed in Scilit:
- Can pulsed xenon ultraviolet light systems disinfect aerobic bacteria in the absence of manual disinfection?American Journal of Infection Control, 2015
- Integrated Chemical and Toxicological Investigation of UV-Chlorine/Chloramine Drinking Water TreatmentEnvironmental Science & Technology, 2014
- UV emissions from low energy artificial light sourcesPhotodermatology, Photoimmunology & Photomedicine, 2014
- Anforderungen an die Hygiene bei der Aufbereitung von MedizinproduktenBundesgesundheitsblatt - Gesundheitsforschung - Gesundheitsschutz, 2012
- Terminal Decontamination of Patient Rooms Using an Automated Mobile UV Light UnitInfection Control & Hospital Epidemiology, 2011
- Endoscopy-related infection: relic of the past?Current Opinion in Infectious Diseases, 2008
- Surface disinfection by exposure to germicidal UV lightIndian Journal of Medical Microbiology, 2008
- High-Level Disinfection or “Sterilization” of Endoscopes?Infection Control & Hospital Epidemiology, 1996
- Anticipatics, handwashing, and handwashing facililtiesAmerican Journal of Infection Control, 1983
- Strahlenchemische Veränderung der Bakterien-Desoxyribonucleinsäure in vivoThe Science of Nature, 1960