New Axis-Marking Method for a Toric Intraocular Lens: Mapping Method
Authors:
- Su Yeon KangSKSu Yeon Kang
- Seung-Hyun KimSKSeung-Hyun Kim
- Jong-Suk SongJSJong-Suk Song
- Hyo-Myung KimHKHyo-Myung Kim
- 1 May 2011
- journal article
- clinical trial
- Published by SLACK, Inc. in Journal of Refractive Surgery
- Vol. 27 (5) , 375-379
Abstract
Purpose: A new axis-marking method, the mapping method, for toric intraocular lens (IOL) implantation is described, and its accuracy is compared with that of conventional methods. Methods: For toric IOL implantation, the steepest corneal axis was determined by IOLMaster optical biometer measurement (Carl Zeiss Meditec). Three marking methods for accurate alignment of the IOL astigmatism axis to the steepest corneal axis were evaluated and compared. For method 1, the patient was seated at the surgical table and instructed to gaze at a distant target. Using a toric reference marker, the corneal limbus was marked at the 3-, 6-, and 9-o’clock positions. The goal axis was also marked using a toric axis marker. Method 2 used a horizontal slit beam for reference marking points at 3 and 9 o’clock. Method 3 was the new mapping method, in which an anterior segment photograph was used to identify several reference vessel points and axis marking points and to calculate actual lengths from the reference vessel points to the axis marking points. During surgery, the axis marking points were marked on the limbus of the eye using calipers. The accuracy of the three axis-marking methods was evaluated using anterior segment photographs. Results: With method 1, the average axis-marking error was 3.69±1.49°. With method 2, the average axis-marking error was 3.14±1.64°. With method 3, the new mapping method, the average axis-marking error was 2.29±1.06°. The new mapping method was more accurate than methods 1 and 2 ( P <.01 and P =.016, respectively, paired t test). Axis-marking errors of the two conventional methods were not significantly different ( P =.061). Conclusions: The new axis-marking method, which uses distinct conjunctival vessels as a reference point, showed less axis-marking error compared to the conventional methods examined in this study.Keywords
This publication has 14 references indexed in Scilit:
- Postoperative corneal shape changes: Microincision versus small-incision coaxial cataract surgeryJournal of Cataract & Refractive Surgery, 2009
- Foldable toric intraocular lens for astigmatism correction in cataract patientsJournal of Cataract & Refractive Surgery, 2008
- Limbal relaxing incisions versus on-axis incisions to reduce corneal astigmatism at the time of cataract surgeryJournal of Cataract & Refractive Surgery, 2005
- Combined toric intraocular lens implantation and relaxing incisions to reduce high preexisting astigmatismJournal of Cataract & Refractive Surgery, 2002
- Toric intraocular lens implantation: 100 consecutive casesJournal of Cataract & Refractive Surgery, 2002
- Toric intraocular lenses for correcting astigmatism in 130 eyesOphthalmology, 2000
- Astigmatism correction with a foldable toric intraocular lens in cataract patientsJournal of Cataract & Refractive Surgery, 2000
- Toric intraocular lenses: Correcting astigmatism while controlling axis shiftJournal of Cataract & Refractive Surgery, 1994
- Cataract patients in a defined Swedish population 1986–1990Acta Ophthalmologica, 1994
- Biometry of 7,500 Cataractous EyesAmerican Journal of Ophthalmology, 1980