Accuracy Assessment of Catheter Guidance Technology in Electrophysiology Procedures
Authors:
- 21 August 2013
- journal article
- research article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 25 (1) , 74-83
Abstract
Accuracy of Catheter Guidance Technology in the EP Lab: Background: With increasing complexity in electrophysiology (EP) procedures, the use of electroanatomic mapping systems (EAMS) as a supplement to fluoroscopy has become common practice. This is the first study that evaluates spatial and point localization accuracy for 2 current EAMS, CARTO3® (Biosense Webster, Diamond Bar, CA, USA) and EnSite Velocity® (St. Jude Medical Inc., St. Paul, MN, USA), and for a novel overlay guidance (OG) software (Siemens AG, Forchheim, Germany) in a phantom experiment.Methods and Results: A C‐arm CT scan was performed on an acrylic phantom containing holes and location markers. Spatial accuracy was assessed for each system using distance measurements involving known markers inside the phantom and properly placed catheters. Anatomical maps of the phantom were acquired by each EAMS, whereas the 3D‐based OG software superimposed an overlay image of the phantom, segmented from the C‐arm CT data set, onto biplane fluoroscopy. Registration processes and landmark measurements quantitatively assessed the spatial accuracy of each technology with respect to the ground truth phantom. Point localization performance was 0.49 ± 0.25 mm in OG, 0.46 ± 0.17 mm in CARTO3® and 0.79 ± 0.83 mm in EnSite®. The registration offset between virtual visualization and reality was 1.10 ± 0.52 mm in OG, 1.62 ± 0.77 mm in CARTO3® and 2.02 ± 1.21 mm in EnSite®. The offset to phantom C‐arm CT landmark measurements was 0.30 ± 0.26 mm in OG, 0.24 ± 0.21 mm in CARTO3® and 1.32 ± 0.98 mm in EnSite®.Conclusions: Each of the evaluated EP guidance systems showed a high level of accuracy; the observed offsets between the virtual 3D visualization and the real phantom were below a clinically relevant threshold of 3 mm.Keywords
Funding Information
- National Institutes of Health (R01HL087917)
- Lucas Foundation
- German Federal Ministry of Education and Research (BMBF) (13EX1012A/E)
- Siemens AG
This publication has 16 references indexed in Scilit:
- Towards cardiac and respiratory motion characterization from electrophysiology data for improved real time MR-integrationJournal of Cardiovascular Magnetic Resonance, 2013
- Pulmonary Vein Isolation Supported by MRI‐Derived 3D‐Augmented Biplane Fluoroscopy: A Feasibility Study and a Quantitative Analysis of the Accuracy of the TechniqueJournal of Cardiovascular Electrophysiology, 2012
- Catheter ablation of atrial fibrillation without the use of fluoroscopyHeart Rhythm, 2010
- Segmentation and registration of three-dimensional rotational angiogram on live fluoroscopy to guide atrial fibrillation ablation: A new online imaging toolHeart Rhythm, 2009
- Validation of Computed Tomography Image Integration into the EnSite NavX Mapping System to Perform Catheter Ablation of Atrial FibrillationJournal of Cardiovascular Electrophysiology, 2008
- Computed tomography-fluoroscopy overlay evaluation during catheter ablation of left atrial arrhythmiaEP Europace, 2008
- Time-resolved three-dimensional imaging of the left atrium and pulmonary veins in the interventional suite—A comparison between multisweep gated rotational three-dimensional reconstructed fluoroscopy and multislice computed tomographyHeart Rhythm, 2008
- Image integration using NavX fusion: Initial experience and validationHeart Rhythm, 2008
- Electrical isolation of pulmonary veins in patients with atrial fibrillation: reduction of fluoroscopy exposure and procedure duration by the use of a non-fluoroscopic navigation system (NavX®)EP Europace, 2006
- A Novel Method for Nonfluoroscopic Catheter-Based Electroanatomical Mapping of the HeartCirculation, 1997