Evaluation of hemorrhagic shock and fluid resuscitation in pigs using handless Doppler carotid artery ultrasound
Authors:
Open Access
- 1 March 2024
- journal article
- research article
- Published by Springer Nature in European Journal of Trauma and Emergency Surgery
- Vol. 50 (4) , 1513-1520
Abstract
Objective: This study aimed to utilize a hemorrhagic shock pig model to compare two hemodynamic monitoring methods, pulse index continuous cardiac output (PiCCO) and spectral carotid artery Doppler ultrasound (CDU). Additionally, we sought to explore the feasibility of employing CDU as a non-invasive hemodynamic monitoring tool in the context of hemorrhagic shock and fluid resuscitation. Design: Animal experiments. Setting and subjects: Female pigs were selected, and hemorrhagic shock was induced by rapid bleeding through an arterial sheath. Interventions: Hemodynamic monitoring was conducted using both PiCCO and CDU during episodes of hemorrhagic shock and fluid resuscitation. Measurements and main results: Among the 10 female pigs studied, CDU measurements revealed a significant decrease in carotid velocity time integral (cVTI) compared to baseline values under shock conditions. During the resuscitation phase, after the mean arterial pressure (MAP) returned to its baseline level, there was no significant difference between cVTI and baseline values. A similar trend was observed for carotid peak velocity (cPV). The corrected flow time (FTc) exhibited a significant difference only at the time of shock compared to baseline values. In comparison to PiCCO, there was a significant correlation between cVTI and MAP (r = 0.616, P < 0.001), stroke volume (SV) (r = 0.821, P < 0.001), and cardiac index (CI) (r = 0.698, P < 0.001). The carotid Doppler shock index (cDSI) displayed negative correlations with MAP (r = − 0.593, P < 0.001), SV (r = − 0.761, P < 0.001), and CI (r = − 0.548, P < 0.001), while showing a positive correlation with the shock index (SI) (r = 0.647, P < 0.001). Conclusions: Compared to PiCCO, CDU monitoring can reliably reflect the volume status of hemorrhagic shock and fluid resuscitation. CDU offers the advantages of being non-invasive, providing real-time data, and being operationally straightforward. These characteristics make it a valuable tool for assessing and managing hemorrhagic shock, especially in resource-limited settings.Keywords
Funding Information
- National Natural Science Foundation of China (82274131)
This publication has 21 references indexed in Scilit:
- Hands-free continuous carotid Doppler ultrasound for detection of the pulse during cardiac arrest in a porcine modelResuscitation Plus, 2023
- Evaluating corrected carotid flow time as a non-invasive parameter for trending cardiac output and stroke volume in cardiac surgery patientsJournal of Ultrasound, 2022
- Carotid Ultrasound to Predict Fluid ResponsivenessJournal of Ultrasound in Medicine, 2020
- Comparing Changes in Carotid Flow Time and Stroke Volume Induced by Passive Leg RaisingThe American Journal of the Medical Sciences, 2018
- Common Carotid Artery Sonography Versus Transthoracic Echocardiography for Cardiac Output Measurements in Intensive Care Unit PatientsJournal of Ultrasound in Medicine, 2017
- A Pilot Assessment of Carotid and Brachial Artery Blood Flow Estimation Using Ultrasound Doppler in Cardiac Surgery PatientsJournal of Cardiothoracic and Vascular Anesthesia, 2016
- Carotid artery corrected flow time measurement via bedside ultrasonography in monitoring volume statusJournal of Critical Care, 2015
- Carotid Flow Time Changes With Volume Status in Acute Blood LossAnnals of Emergency Medicine, 2015
- World Medical Association Declaration of HelsinkiJAMA, 2013
- Global and regional mortality from 235 causes of death for 20 age groups in 1990 and 2010: a systematic analysis for the Global Burden of Disease Study 2010The Lancet, 2012