Application of a micro-droplet generator for an ICP-sector field mass spectrometer – optimization and analytical characterization
- 4 January 2013
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 28 (5) , 646-656
Abstract
A micro-droplet generator (μDG) sample introduction system was coupled to a sector field ICP-MS instrument to investigate the analytical figures of merit with respect to single cell analysis. The sector field instrument was operated for the first time in a fast scanning mode (E-scan) with the shortest time resolution of 100 μs to measure the single droplet time resolved and using the original detector in a pulse counting mode without modification of the existing electronics. For reduction of the droplet diameter a triple pulse mode of the droplet generator was applied and a droplet diameter down to 23 μm has been achieved for this investigation with a 100% transport efficiency of droplets. Signal duration times of single droplets of less than 500 μs have been measured. Overall detection efficiencies in the range of 10−3 counts per atom have been achieved and absolute limits of detection range between 120 ag for Fe and 1.1 ag for Mg as a mean value from 1000 droplet events.Keywords
This publication has 22 references indexed in Scilit:
- Quantitative Imaging of Gold and Silver Nanoparticles in Single Eukaryotic Cells by Laser Ablation ICP-MSAnalytical Chemistry, 2012
- Considerations for measurement of individual nanoparticles or microparticles by ICP-MS: determination of the number of particles and the analyte mass in each particleJournal of Analytical Atomic Spectrometry, 2012
- LA-ICP-MS and nHPLC-ESI-LTQ-FT-MS/MS for the analysis of cisplatin–protein complexes separated by two dimensional gel electrophoresis in biological samplesJournal of Analytical Atomic Spectrometry, 2012
- Multiplexed Immunohistochemical Detection of Tumor Markers in Breast Cancer Tissue Using Laser Ablation Inductively Coupled Plasma Mass SpectrometryAnalytical Chemistry, 2011
- Iodine as an elemental marker for imaging of single cells and tissue sections by laser ablation inductively coupled plasma mass spectrometryJournal of Analytical Atomic Spectrometry, 2011
- Elemental Bioimaging in Kidney by LA–ICP–MS As a Tool to Study Nephrotoxicity and Renal Protective Strategies in Cisplatin TherapiesAnalytical Chemistry, 2011
- Application of plasma spectrometry for the analysis of engineered nanoparticles in suspensions and productsJournal of Analytical Atomic Spectrometry, 2011
- Chemical Analysis of Single CellsAnalytical Chemistry, 2011
- Time-resolved ICP-MS measurement for single-cell analysis and on-line cytometryJournal of Analytical Atomic Spectrometry, 2010
- Metallomics study on all-elements analysis of salmon egg cells and fractionation analysis of metals in cell cytoplasmPure and Applied Chemistry, 2008