Greater accuracy in positioning of the acetabular cup by using an image-free navigation system

Authors:

Abstract
In a prospective and randomised clinical study, acetabular cups were implanted free-hand (control group n=22) or with computer assistance using an image-free navigation system (study group n=23). The cup position was determined postoperatively on pelvic CT. An average inclination of 42.3° (range: 30°–53°; SD±7.0°) and an average anteversion of 24.0° (range: −3° to 51°; SD±15.0°) were found in the control group, and an average inclination of 45.0° (ranage: 40°–50°; SD±2.8°) and an average anteversion of 14.4° (range: 5°–25°; SS±5.0°) in the computer-assisted study group. The deviations from the desired cup position (45° inclination, 15° anteversion) were significantly lower in the computer-assisted study group (pp=0.003). Dans un étude clinique prospective et randomisé, les cupules acétabulaires ont été implantées de façon habituelle (n=22; groupe témoin) ou avec assistance d’un ordinateur qui utilise un système de navigation image—libre (n=23; groupe d’étude). La place de la cupule a été déterminée après l’opération sur un scanner pelvien. Une inclinaison moyenne de 42,3° (30° à 53°; ±7.0°) et une antéversion moyenne de 24,0° (−3° à 51°; ±15.0°) ont été trouvées dans le groupe témoin et une inclinaison moyenne de 45,0° (40° à 50°; ±2.8°) et une antéversion moyenne de 14,4° (5° à 25°; ±5.0°) dans le groupe de l’étude assistée par ordinateur. Les déviations par rapport à la position désirée de la cupule (45° d’inclinaison, 15° d’antéversion) étaient notablement inférieures dans le groupe de l’étude assistée par ordinateur (pp=0.003).