Screening electro-oculography protocol as a part of full-field electroretinography
- 15 November 2024
- journal article
- research article
- Published by Springer Nature in Documenta Ophthalmologica
- Vol. 149 (3) , 143-150
Abstract
Purpose To simplify the electro-oculography (EOG) method and integrate it into the full-field electroretinogram (ffERG) protocol for screening purposes. Methods 20 control subjects and 5 patients with Best vitelliform macular dystrophy (BVMD) underwent EOG recording according to both the standard protocol and screening EOG protocol that was integrated as part of ffERG testing. Mean values of light peak-to-dark trough ratio (LP:DT ratio) were compared between both protocols using the Student’s t-test, sensitivity and specificity for the detection of RPE dysfunction were evaluated with ROC analysis and a survey on the difficulty of each protocol was completed by each subject. Results With the standard EOG mean LP:DT ratio was 2.67 ± 0.61 in controls and 1.12 ± 0.16 in BVMD patients (p < 0.001). With the screening protocol mean LP:DT ratio was 1.98 ± 0.33 in controls and 1.02 ± 0.14 in BVMD (p < 0.001). A comparison of LP:DT ratios showed significant difference between the standard and the screening protocol (p < 0.01 in controls; p = 0.02 in BVMD), however both protocols showed 100% sensitivity and 100% specificity for detection of BVMD. Patients stated that participation in the screening protocol was easier and less uncomfortable. Conclusions Screening EOG performed as part of ffERG gives comparable results to standard EOG, examination is patient-friendly, time saving and can be used as a preliminary test for the assessment of RPE function.Keywords
This publication has 16 references indexed in Scilit:
- ISCEV Standard for full-field clinical electroretinography (2022 update)Documenta Ophthalmologica, 2022
- A meta-analysis of clinical electro-oculography valuesDocumenta Ophthalmologica, 2017
- ISCEV Standard for clinical electro-oculography (2017 update)Documenta Ophthalmologica, 2017
- Biallelic Mutation of BEST1 Causes a Distinct Retinopathy in HumansThe American Journal of Human Genetics, 2008
- The electro-oculogramProgress in Retinal and Eye Research, 2006
- Electro-oculography in Autosomal Dominant VitreoretinochoroidopathyArchives of Ophthalmology (1950), 1992
- Fast and Slow Oscillations of the Electro-oculogram in Best's Macular Dystrophy and Retinitis PigmentosaArchives of Ophthalmology (1950), 1989
- Electro-Oculography in Best's Macular DystrophyAmerican Journal of Ophthalmology, 1974
- Electro-oculography in Vitelliform Degeneration of the MaculaArchives of Ophthalmology (1950), 1967
- Hereditary Vitelliruptive Macular DegenerationAmerican Journal of Ophthalmology, 1966