First multicenter experience using the Silk Vista flow diverter in 60 consecutive intracranial aneurysms: technical aspects
Open Access
- 8 April 2021
- journal article
- research article
- Published by BMJ in Journal of NeuroInterventional Surgery
- Vol. 13 (12) , 1145-1151
Abstract
Background The aim of this study was to assess the technical success and procedural safety of the new Silk Vista device (SV) by evaluating the intraprocedural and periprocedural complication rate after its use in several institutions worldwide. Methods The study involved a retrospective review of multicenter data regarding a consecutive series of patients with intracranial aneurysms, treated with the SV between September 2020 and January 2021. Clinical, intra/periprocedural and angiographic data, including approach, materials used, aneurysm size and location, device/s, technical details and initial angiographic aneurysm occlusion, were analyzed. Results 60 aneurysms were treated with SV in 57 procedures. 66 devices were used, 3 removed and 63 implanted. The devices opened instantaneously in 60 out of 66 (91%) cases and complete wall apposition was achieved in 58 out of 63 (92%) devices implanted. In 4 out of 66 (6%) devices a partial opening of the distal end occurred, and in 5 (8%) devices incomplete apposition was reported. There were 3 (5%) intraprocedural thromboembolic events managed successfully with no permanent neurological morbidity, and 4 (7%) postprocedural events. There was no mortality in this study. The initial occlusion rates in the 60 aneurysms were as follows: O’Kelly–Marotta (OKM) A in 34 (57%) cases, OKM B in 15 (25%) cases, OKM C in 6 (10%) cases, and OKM D in 5 (8%) cases. Conclusions Our study demonstrated that the use of the new flow diverter Silk Vista for the treatment of intracranial aneurysms is feasible and technically safe.Keywords
This publication has 27 references indexed in Scilit:
- Transient in-stent stenosis at mid-term angiographic follow-up in patients treated with SILK flow diverter stents: incidence, clinical significance and long-term follow-upJournal of NeuroInterventional Surgery, 2018
- Polymer coating embolism from intravascular medical devices — a clinical literature reviewCardiovascular Pathology, 2017
- Republished: The bumper technique for advancing a large profile microcatheterJournal of NeuroInterventional Surgery, 2017
- Utilization of a Novel, Multi-Durometer Intracranial Distal Access Catheter: Nuances and Experience in 110 Consecutive Cases of Aneurysm Flow DiversionInterventional Neurology, 2017
- Use of self-expanding stents for better intracranial flow diverter wall appositionInterventional Neuroradiology, 2016
- Pentaxial access platform for ultra-distal intracranial delivery of a large-bore hyperflexible DIC (distal intracranial catheter): A technical noteInterdisciplinary Neurosurgery, 2016
- Flow-diverter devices in the treatment of intracranial aneurysms: A meta-analysis and systematic reviewThe Neuroradiology Journal, 2016
- Silk Flow-Diverter Stent for the Treatment of Intracranial Aneurysms: A Series of 58 Patients with Emphasis on Long-Term ResultsAmerican Journal of Neuroradiology, 2014
- Thromboembolic complications with Pipeline Embolization Device placement: impact of procedure time, number of stents and pre-procedure P2Y12 reaction unit (PRU) valueJournal of NeuroInterventional Surgery, 2014
- The Pipeline Embolization DeviceNeurosurgery, 2013