Audiometer calibration: interpreting and applying the standards.
- 1 October 2000
- journal article
- research article
- Published by Taylor & Francis in British Journal of Audiology
- Vol. 34 (5) , 311-316
Abstract
A degree of judgement is required when applying the British Standard associated with the calibration and performance of audiometers, BS EN 60645-1 (IEC 645-1). Consequently there is potential for variation of practice between calibration agencies. This note addresses issues and makes suggestions relating to the measurement of distortion, on-off effect, cross talk, linearity, rise/fall time and to the positioning of the earphone on an artificial ear. The degree of precision appropriate to each measurement is suggested. A warning is given of the audiological consequences of excessive distortion produced by normally functioning audiometers.Keywords
This publication has 3 references indexed in Scilit:
- Audiometer calibration: two neglected areas.British Journal of Audiology, 1997
- Audiometer calibration: applying the earphone to the couplerBritish Journal of Audiology, 1995
- Bone conduction errors at high frequencies: implications for clinical and medico-legal practiceThe Journal of Laryngology & Otology, 1993