The evaluation of low cut‐off index values of Elecsys® HIV combi PT assay in predicting false‐positive results
Open Access
- 25 August 2020
- journal article
- research article
- Published by Wiley in Journal of Clinical Laboratory Analysis
- Vol. 34 (11)
Abstract
Objective: To analyze the results of different cut‐off index (COI) values of Elecsys® HIV combi PT assay and to assess the role of COI in reducing the frequency of false‐positive results.Methods: We conducted a retrospective study of samples analyzed by Elecsys® HIV combi PT assay, a 4th‐generation ECLIA, between 2016 and 2017. A total amount of 379 122 samples were collected for HIV (Human Immunodeficiency Virus) screening.Results: A total of 379 122 samples were analyzed. 2528 (0.67%) were positive by Elecsys® HIV combi PT. Of these, 468 were false‐positive results, and most of them (94.87%) were in samples with 1 < COI < 15. The false‐positive rate was 0.12%. Patients with false‐positive samples were more distributed in elder (P < .001) and female (P < .001) than true‐positive specimens. The median COI in true‐positive specimens was (385.20), which is significantly higher than false‐positive specimens (2.08). The consistency between Elecsys® HIV combi PT assay and 3rd‐generation and positive predictive value (PPV) increased with higher COI values. Cancer, infection, and neurological diseases were considered the potential confounding factors of HIV false‐positive results (19.44%, 11.11%, and 6.62%, respectively).Conclusion: Samples with low COI values, especially those contain confounding factors, need to be further scrutinized to determine whether the confounding factors may cause false‐positive problem. In addition, the hypothesis that low COI values may predict false‐positive results is valid.Keywords
This publication has 20 references indexed in Scilit:
- Human Immunodeficiency Virus Diagnostic Testing: 30 Years of EvolutionClinical and Vaccine Immunology, 2016
- Spectrum of false positivity for the fourth generation human immunodeficiency virus diagnostic testsAIDS Research and Therapy, 2016
- Accounting for False Positive HIV Tests: Is Visceral Leishmaniasis Responsible?PLOS ONE, 2015
- Systemic lupus erythematosus patient with false positive results of antibody to HIV: A case report and a comprehensive literature reviewTechnology and Health Care, 2015
- Clinical Application Evaluation of Two Fourth-Generation Human Immunodeficiency Virus (HIV) Screening Assays in West China HospitalJournal of Clinical Laboratory Analysis, 2014
- Performance evaluation of a new fourth-generation HIV Ag/Ab combination electrochemiluminescence immunoassay – evaluation of a new HIV assayInternational Journal of STD & AIDS, 2013
- Performances of four fourth‐generation human immunodeficiency virus‐1 screening assaysJournal of Medical Virology, 2012
- Performance evaluation of a new fourth-generation HIV combination antigen–antibody assayMedical Microbiology and Immunology, 2012
- Interpretation of HIV Serologic Testing ResultsClinical Chemistry, 2010
- A Human Immunodeficiency Virus Screening Algorithm to Address the High Rate of False-Positive Results in Pregnant Women in JapanPLOS ONE, 2010