Evaluation of a planar diode matrix for SRS patient‐specific QA in comparison with GAFchromic films
Open Access
- 5 July 2023
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 24 (8) , e13947
Abstract
Purpose: We validate the routine use of a two‐dimensional (2D) diode matrix for patient specific pre‐treatment verification for Cyberknife (CK) stereotactic radiosurgery and to compare it with film dosimetry.Materials and method: A total of 46 patients were selected according to the most frequent diseases treated at our institution with the CK system, that is, brain metastases, meningiomas, spine metastases, and prostate tumors. All cases were evaluated with GAFChromic EBT‐3 films and SRS MapCHECK for Fixed cone, IRIS, and MLC collimators of the CK.Results: The highest mean passing rate was observed for the SRS MapCHECK system compared to films. In order to assess if the two techniques provide statistically different results, a Wilcoxon Signed‐Rank non‐parametric test was performed (p < 0.05) and we found gamma values significantly lower for EBT‐3 films with respect to the SRS MapCHECK. We noticed a moderately significant association between the two techniques using Spearman's rank correlation coefficient (rs > 0.4). We also performed the Bland–Altman statistical method: less than 5% of the differences resulted outside the range (mean ± 1.96 × SD), so the two methods can be considered interchangeable within the combined inaccuracy.Conclusions: The use of SRS MapCHECK for CK patient specific quality assurance (QA) is feasible for a variety of clinical districts and could be reliably used as a replacement for radiochromic films.Keywords
This publication has 33 references indexed in Scilit:
- Penalization of aperture complexity in inversely planned volumetric modulated arc therapyMedical Physics, 2012
- Assessing software upgrades, plan properties and patient geometry using intensity modulated radiation therapy (IMRT) complexity metricsMedical Physics, 2011
- Stereotactic body radiation therapy: The report of AAPM Task Group 101Medical Physics, 2010
- Evaluation of the Delta4phantom for IMRT and VMAT verificationPhysics in Medicine & Biology, 2009
- The Use of TLD and Gafchromic Film to Assure Submillimeter Accuracy for Image-Guided RadiosurgeryMedical Dosimetry, 2008
- Spatial resolution of 2D ionization chamber arrays for IMRT dose verification: single-detector size and sampling step widthPhysics in Medicine & Biology, 2007
- Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routinePhysics in Medicine & Biology, 2007
- Characterization of a 2D ion chamber array for the verification of radiotherapy treatmentsPhysics in Medicine & Biology, 2005
- CyberKnife®: A Robotic Radiosurgery SystemClinical Journal of Oncology Nursing, 2002
- The Cyberknife: A Frameless Robotic System for RadiosurgeryStereotactic and Functional Neurosurgery, 1997