Two-dimensional position-sensitive spectrometer for registration of ionizing radiation
Authors:
Open Access
- 1 November 2019
- conference paper
- Published by IOP Publishing in Journal of Physics: Conference Series
- Vol. 1400 (5) , 055050
Abstract
A new two-coordinate position-sensitive spectrometer for the measurements of spatial and energy distribution of ionizing particles was developed. The sensitive elements of the device are two orthogonal 64 strip semiconductor detectors whose signals are read out by two Application-Specific Integrated Circuits (ASIC) IDE1140 comprising 64 spectrometric channels each. Control, pre-processing data acquisition and transfer to a computer are carried out using FPGA circuit EP3C16Q240. The main spectrometer characteristics obtained in the measurements are: equivalent noise charge of 0.7 fC and integral nonlinearity is below 1% in the range from 5 to 160 fC.Keywords