A semi-automated natural language processing-based analytical pipeline for identifying structural shifts in MRI safety incidents: an analysis of FDA MAUDE reports (2018–2024)
Authors:
- Jeongin LeeJLJeongin Lee
- Seongdong ParkSPSeongdong Park
- Jiwon JungJJJiwon Jung
- Jinwoo HwangJHJinwoo Hwang
- Seon-Eui HongSHSeon-Eui Hong
- Yeun-Chul RyuYRYeun-Chul Ryu
- Myung Kyun WooMWMyung Kyun Woo
- Kisoo KimKKKisoo Kim
- Sukhoon OhSOSukhoon Oh
Open Access
- 16 June 2026
- journal article
- Published by Springer Nature in Journal of Analytical Science and Technology
- Vol. 17 (1)
Abstract
Despite the widespread adoption of magnetic resonance imaging (MRI), safety incidents during MRI examinations continue to be reported. Building upon prior FDA-led analyses of MAUDE data, this study extends the temporal scope and analytical depth to systematically examine MRI-related adverse events reported between 2018 and 2024. Using a semi-automated classification framework that combined keyword-based text analysis with expert review, MRI-related incidents were categorized into granular safety domains, and inter-category correlations and associations were analyzed to characterize incident patterns at the level of individual events. The results demonstrate that MRI-related incidents persist annually, with no substantial reduction observed in Operation- and RF/Gradient-related events, which together account for a large proportion of reported cases. In parallel, implant-related incidents have become increasingly prominent and are associated with the majority of reported MRI events, extending beyond any single failure category. Although further subclassification of implant-related incidents was attempted, substantial semantic overlap among incident descriptions highlighted the complex and intertwined nature of implant-associated MRI safety challenges. Overall, these findings indicate that contemporary MRI safety risks are increasingly shaped by interactions among implants, software-based systems, and operational factors, rather than isolated hardware failures alone. Furthermore, the developed analytical pipeline establishes a reproducible measurement framework for large-scale medical device surveillance. This study provides data-driven insights that may inform future medical device design, more conservative MRI safety guidelines, and expanded education for both medical professionals and patients.Keywords
Funding Information
- Kyung Hee University (GS-1-JO-ON- 20252365)
This publication has 18 references indexed in Scilit:
- Recent Advancements in Cardiac Implantable Devices for Pediatric PatientsCurrent Pediatrics Reports, 2024
- A narrative review of current and emerging MRI safety issues: What every MRI technologist (radiographer) needs to knowJournal of Medical Radiation Sciences, 2021
- Unexpected magnetic resonance imaging burn injuries from jogging pantsRadiology Case Reports, 2019
- MRI‐related FDA adverse event reports: A 10‐yr reviewMedical Physics, 2019
- BioBERT: a pre-trained biomedical language representation model for biomedical text miningBioinformatics, 2019
- Magnetic Resonance Imaging of Soft Tissue MassesSeminars in Roentgenology, 2017
- Multiparametric imaging with heterogeneous radiofrequency fieldsNature Communications, 2016
- Interestingness measures for association rules based on statistical validityKnowledge-Based Systems, 2011
- Investigation of the factors responsible for burns during MRIJournal of Magnetic Resonance Imaging, 2001
- Safety of Strong, Static Magnetic FieldsJournal of Magnetic Resonance Imaging, 2000