BACKGROUND: Prospective studies have established the safety and efficacy of the PipelineTM Embolization Device (PED; Medtronic) for treatment of intracranial aneurysms (IA). OBJECTIVE: To investigate long-term outcomes from the Pipeline Embolization Devices for the Treatment of Intracranial Aneurysms (PEDESTRIAN) Registry. METHODS: The PEDESTRIAN Registry data were retrospectively reviewed, which included patients (March 2006 to July 2019) with complex IAs treated with PED. Patients with unfavorable anatomy and/or recurrence following previous treatment were included and excluded those with acute subarachnoid hemorrhage. The primary angiographic endpoint was complete occlusion and long-term stability. Clinical and radiological follow-up was performed at 3 to 6 mo, 12 mo, and yearly thereafter. RESULTS: A total of 835 patients (mean age 55.9 ± 14.7 yr; 80.0% female) with 1000 aneurysms were included. Aneurysms varied in size: 64.6% were small (≤10 mm), 25.6% were large (11-24 mm), and 9.8% were giant (≥25 mm). A total of 1214 PEDs were deployed. Follow-up angiography was available for 85.1% of patients with 776 aneurysms at 24.6 ± 25.0 mo (mean). Complete occlusion was demonstrated in 75.8% of aneurysms at 12 mo, 92.9% at 2 to 4 yr, and 96.4% at >5 yr. During the postprocedural period, modified Rankin Scale scores remained stable or improved in 96.2% of patients, with stability or improvement in 99.1% of patients >5 yr. The overall major morbidity and neurological mortality rate was 5.8%. CONCLUSION: This study demonstrated high rates of long-term complete aneurysm occlusion, stable or improved functional outcomes, and low rates of complications and mortality. Clinical and angiographic outcomes improved over long-term follow-up, demonstrating that endovascular treatment of IA with PED is safe and effective.Keywords
This publication has 17 references indexed in Scilit:
- Safety and efficacy of the Pipeline embolization device for treatment of intracranial aneurysms: a pooled analysis of 3 large studiesJournal of Neurosurgery, 2017
- Treatment of Distal Anterior Circulation Aneurysms With the Pipeline Embolization DeviceNeurosurgery, 2016
- Aneurysm Study of Pipeline in an Observational Registry (ASPIRe)Interventional Neurology, 2016
- Safety and efficacy of the Pipeline Embolization Device in 100 small intracranial aneurysmsJournal of Neurosurgery, 2015
- International Retrospective Study of the Pipeline Embolization Device: A Multicenter Aneurysm Treatment StudyAmerican Journal of Neuroradiology, 2014
- Cellular Mechanisms of Aneurysm Occlusion after Treatment with a Flow DiverterRadiology, 2014
- Pipeline for Uncoilable or Failed Aneurysms: Results from a Multicenter Clinical TrialRadiology, 2013
- Intracranial Aneurysms: Midterm Outcome of Pipeline Embolization Device—A Prospective Study in 143 Patients with 178 AneurysmsRadiology, 2012
- The Pipeline Embolization Device for the Intracranial Treatment of Aneurysms TrialAmerican Journal of Neuroradiology, 2010
- Endovascular Treatment of Unruptured AneurysmsStroke, 2001