Simultaneous multislice imaging for native myocardial T1 mapping: Improved spatial coverage in a single breath-hold
- 5 June 2017
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 78 (2) , 462-471
Abstract
Purpose To develop a saturation recovery myocardial T1 mapping method for the simultaneous multislice acquisition of three slices. Methods Saturation pulse‐prepared heart rate independent inversion recovery (SAPPHIRE) T1 mapping was implemented with simultaneous multislice imaging using FLASH readouts for faster coverage of the myocardium. Controlled aliasing in parallel imaging (CAIPI) was used to achieve minimal noise amplification in three slices. Multiband reconstruction was performed using three linear reconstruction methods: Slice‐ and in‐plane GRAPPA, CG‐SENSE, and Tikhonov‐regularized CG‐SENSE. Accuracy, spatial variability, and interslice leakage were compared with single‐band T1 mapping in a phantom and in six healthy subjects. Results Multiband phantom T1 times showed good agreement with single‐band T1 mapping for all three reconstruction methods (normalized root mean square error 1 times showed no significant difference compared with single‐band (T1 time ± intersegmental variability: single‐band, 1580 ± 119 ms; GRAPPA, 1572 ± 145 ms; CG‐SENSE, 1579 ± 159 ms; Tikhonov, 1586 ± 150 ms [analysis of variance; P = 0.86]). Interslice leakage was smallest for GRAPPA (5.4%) and higher for CG‐SENSE (6.2%) and Tikhonov‐regularized CG‐SENSE (7.9%). Conclusion Multiband accelerated myocardial T1 mapping demonstrated the potential for single–breath‐hold T1 quantification in 16 American Heart Association segments over three slices. A 1.2‐ to 1.4‐fold higher in vivo spatial variability was observed, where GRAPPA‐based reconstruction showed the highest homogeneity and the least interslice leakage. Magn Reson Med 78:462–471, 2017. © 2017 International Society for Magnetic Resonance in MedicineKeywords
Funding Information
- National Institutes of Health (R00HL111410, P41EB015894)
This publication has 57 references indexed in Scilit:
- Extracellular volume fraction mapping in the myocardium, part 1: evaluation of an automated methodJournal of Cardiovascular Magnetic Resonance, 2012
- Blipped‐controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging with reduced g‐factor penaltyMagnetic Resonance in Medicine, 2011
- Shortened Modified Look-Locker Inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 1.5 and 3 T within a 9 heartbeat breathholdJournal of Cardiovascular Magnetic Resonance, 2010
- T2 quantification for improved detection of myocardial edemaJournal of Cardiovascular Magnetic Resonance, 2009
- T2 measurement of the human myocardium using a T2‐prepared transient‐state trueFISP sequenceMagnetic Resonance in Medicine, 2007
- Modified Look‐Locker inversion recovery (MOLLI) for high‐resolution T1 mapping of the heartMagnetic Resonance in Medicine, 2004
- Generalized autocalibrating partially parallel acquisitions (GRAPPA)Magnetic Resonance in Medicine, 2002
- Advances in sensitivity encoding with arbitrary k‐space trajectoriesMagnetic Resonance in Medicine, 2001
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- Multifrequency selective rf pulses for multislice MR imagingMagnetic Resonance in Medicine, 1988