Retinal safety of the irradiation delivered to light-adjustable intraocular lenses evaluated in a rabbit model
Authors:
- Liliana WernerLWLiliana Werner
- Wellington ChangWCWellington Chang
- Jonathan HaymoreJHJonathan Haymore
- Brian HaugenBHBrian Haugen
- Natalya RomanivNRNatalya Romaniv
- Christian SandstedtCSChristian Sandstedt
- Shiao ChangSCShiao Chang
- Nick MamalisNMNick Mamalis
- 1 August 2010
- journal article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 36 (8) , 1392-1397
Abstract
To evaluate the safety to the retina of a light-delivery device used to irradiate a light-adjustable intraocular lens (IOL) after implantation in a rabbit model. John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA. In this study, rabbits had phacoemulsification with implantation of an ultraviolet (UV)-filtering light-adjustable IOL (study IOL) in 1 eye and a custom-made silicone IOL without a UV filter (control IOL) in the opposite eye. The study IOLs were irradiated at 1.0, 2.0, 3.0, and 5.0 times the expected maximum UV irradiation doses and the control IOLs, at 0.3, 0.6, 1.0, and 2.0 times. One week after irradiation, slitlamp and fundus (indirect ophthalmoscopy) examinations were performed. The rabbits were then humanely killed and their eyes enucleated and processed for histopathology. The 16 eyes with the study IOL (with UV filter) showed no signs of corneal, anterior segment, or retinal toxicity on histopathologic evaluation. The 16 eyes with the control IOL (no UV filter) also showed no signs of corneal or anterior segment toxicity; however, 3 eyes receiving the higher radiation doses had focal areas of retinal damage consistent with laser burn. Pigmented rabbit eyes with a light-adjustable IOL with a UV filter showed no signs of retina toxicity after near-UV light exposure up to 5 times the expected maximum treatment dosage. No author has a financial or proprietary interest in any material or method mentioned. Additional disclosures are found in the footnotes.Keywords
This publication has 17 references indexed in Scilit:
- Correction of Myopia after Cataract Surgery with a Light-Adjustable LensOphthalmology, 2009
- Correction of Residual Hyperopia After Cataract Surgery Using the Light Adjustable Intraocular Lens TechnologyAmerican Journal of Ophthalmology, 2009
- Complications of foldable intraocular lenses requiring explantation or secondary intervention—2007 survey updateJournal of Cataract & Refractive Surgery, 2008
- Corneal endothelial safety with the irradiation system for light-adjustable intraocular lensesJournal of Cataract & Refractive Surgery, 2007
- Phototoxicity to the Retina: Mechanisms of DamageInternational Journal of Toxicology, 2002
- Spatial Mapping of Polarized Light Transmission in the Central Rabbit CorneaExperimental Eye Research, 1996
- UV-absorbing intraocular lenses: Safety, efficacy, and consequences for the cataract patientGraefe's Archive for Clinical and Experimental Ophthalmology, 1989
- Ultraviolet light absorption in intraocular lensesJournal of Cataract & Refractive Surgery, 1986
- Effect of an Ultraviolet-filtering Intraocular Lens on Cystoid Macular EdemaOphthalmology, 1985
- A schematic eye for the rabbitVision Research, 1972