In Situ Drift Monitoring and Calibration of Field-Deployed Potentiometric Sensors Using Temperature Supervision
Authors:
- 23 June 2023
- journal article
- research article
- Published by American Chemical Society (ACS) in ACS Sensors
- Vol. 8 (7) , 2799-2808
Abstract
Potentiometric ion-selective electrodes (ISEs) have broad applications in personalized healthcare, smart agriculture, oil/gas exploration, and environmental monitoring. However, high-precision potentiometric sensing is difficult with field-deployed sensors due to time-dependent voltage drift and the need for frequent calibration. In the laboratory setting, these issues are resolved by repeated calibration by measuring the voltage response at multiple standard solutions at a constant temperature. For field-deployed sensors, it is difficult to frequently interrupt operation and recalibrate with standard solutions. Moreover, the constant surrounding temperature constraint imposed by the traditional calibration process makes it unsuitable for temperature-varying field use. To address the challenges of traditional calibration for field-deployed sensors, in this study, we propose a novel in situ calibration approach in which we use natural/external temperature variation in the field to obtain the time-varying calibration parameters, without having to relocate the sensors or use any complex system. We also develop a temperature-supervised monitoring method to detect the drift of the sensor during operation. Collectively, the temperature-based drift monitoring and in situ calibration methods allow us to monitor the drift of sensors and correct them periodically to achieve high-precision sensing. We demonstrate our approach in three testbeds: (1) under controlled temperature variation in the lab, (2) under natural temperature variation in a greenhouse, and (3) in the field to monitor nitrate activity of an agricultural site. In the laboratory study, we validate that the calibration parameters of printed nitrate ISEs can be reproduced by our proposed calibration process; therefore, it can serve as an alternative to traditional calibration processes. In the greenhouse, we show the use of natural temperature variation to calibrate the sensors and detect the drift in a fixed concentration nitrate solution. Finally, we demonstrate the use of the method to monitor the nitrate activity of an agricultural field within 10% of laboratory-based measurements (i.e., a sensitivity of 0.03 mM) for a period of 22 days. The findings highlight the prospect of temperature-based calibration and drift monitoring for high-precision sensing with field-deployed ISEs.Keywords
Funding Information
- Wabash Heartland Innovation Network
- Purdue Scalable Manufacturing of Aware and Responsive Thin Films
This publication has 29 references indexed in Scilit:
- Instrument-Free Control of the Standard Potential of Potentiometric Solid-Contact Ion-Selective Electrodes by Short-Circuiting with a Conventional Reference ElectrodeAnalytical Chemistry, 2014
- Electrochemical control of the standard potential of solid-contact ion-selective electrodes having a conducting polymer as ion-to-electron transducerElectrochimica Acta, 2014
- Polymeric plasticizer extends the lifetime of PVC-membrane ion-selective electrodesThe Analyst, 2013
- The influence of an intermediate layer on the composition stability of a polymeric ion-selective membraneElectrochimica Acta, 2012
- Determination of the leaching of polymeric ion-selective membrane components by stripping voltammetryTalanta, 2010
- Light sensitivity and potential stability of electrically conducting polymers commonly used in solid contact ion-selective electrodesJournal of Solid State Electrochemistry, 2008
- Beyond potentiometry: Robust electrochemical ion sensor concepts in view of remote chemical sensingTalanta, 2008
- Estimating the soil temperature profile from a single depth observation: A simple empirical heatflow solutionWater Resources Research, 2008
- A novel naphazoline-selective membrane sensor and its pharmaceutical applicationsSensors and Actuators B: Chemical, 2006
- High-resolution determination of the pH of seawater with a flow-through systemDeep Sea Research Part A. Oceanographic Research Papers, 1988