Real-Time Pachymetry During Photorefractive Keratectomy Using Optical Low-Coherence Reflectometry

Authors:
  • MBMatthias BöhNke
    Matthias BöhNke

Abstract
Optical low-coherence reflectometry (OLCR) was used as a noncontact method to measure the central corneal thickness of three patients intraoperatively during photorefractive keratectomy. Continuous on-line measurements were performed on the intact cornea immediately before the beginning of surgery, after the removal of the corneal epithelium, during laser tissue photoablation, and for 3 min after the ablation process. Corneal thinning due to evaporation was studied on a separate patient with the OLCR instrument, and it was found to be −0.14 mm/s during the first 5 min after epithelium removal. This baseline corneal thinning rate was used as a fit parameter to calculate actual from measured ablation depths. The measurements showed a maximum difference of 610 mm between planned ablations (34–92 mm) and measured ablation depths.