A Comprehensive Method for Calculating Patient Effective Dose and Other Dosimetric Quantities From CT DICOM Images

Authors:

Abstract
OBJECTIVE. The purpose of this article is to present a method for the calculation of effective dose using the DICOM header information of CT images. MATERIALS AND METHODS. Using specialized software, the DICOM data were automatically extracted into a spreadsheet containing embedded functions for calculating effective dose. These data were used to calculate the dose-length product (DLP) fraction that corresponds to each image, and the respective effective dose was obtained by multiplying the image DLP by a conversion coefficient that was automatically selected depending on the CT scanner, the tube potential, and the anatomic position to which each image corresponded. The total effective dose was calculated as the sum of effective doses of all images plus the contribution of overscan. The conversion coefficient tables were derived using dosimetry calculator software for both the International Commission on Radiological Protection (ICRP) 60 and ICRP 103 organ-weighting schemes. This method was applied for 90...