A Comprehensive Method for Calculating Patient Effective Dose and Other Dosimetric Quantities From CT DICOM Images
- 1 July 2012
- journal article
- Published by American Roentgen Ray Society in American Journal of Roentgenology
- Vol. 199 (1) , 133-141
Abstract
OBJECTIVE. The purpose of this article is to present a method for the calculation of effective dose using the DICOM header information of CT images. MATERIALS AND METHODS. Using specialized software, the DICOM data were automatically extracted into a spreadsheet containing embedded functions for calculating effective dose. These data were used to calculate the dose-length product (DLP) fraction that corresponds to each image, and the respective effective dose was obtained by multiplying the image DLP by a conversion coefficient that was automatically selected depending on the CT scanner, the tube potential, and the anatomic position to which each image corresponded. The total effective dose was calculated as the sum of effective doses of all images plus the contribution of overscan. The conversion coefficient tables were derived using dosimetry calculator software for both the International Commission on Radiological Protection (ICRP) 60 and ICRP 103 organ-weighting schemes. This method was applied for 90...Keywords
This publication has 13 references indexed in Scilit:
- The impact of overscan on patient dose with first generation multislice CT scannersPhysica Medica, 2011
- A method for calculating the dose length product from CT DICOM imagesBritish Journal of Radiology, 2011
- Computing effective dose in cardiac CTPhysics in Medicine & Biology, 2010
- Converting Dose-Length Product to Effective Dose at CTRadiology, 2008
- Radiation Dose Modulation Techniques in the Multidetector CT Era: From Basics to PracticeRadioGraphics, 2008
- CT-Guided Interventional Procedures without CT Fluoroscopy Assistance: Patient Effective Dose and Absorbed Dose ConsiderationsAmerican Journal of Roentgenology, 2007
- Overranging in Multisection CT: Quantification and Relative Contribution to Dose—Comparison of Four 16-Section CT ScannersRadiology, 2007
- CT Dose Reduction and Dose Management Tools: Overview of Available OptionsRadioGraphics, 2006
- The effect of overscanning on patient effective dose from multidetector helical computed tomography examinationsMedical Physics, 2005
- Primary radiation outside the imaged volume of a multislice helical CT scanBritish Journal of Radiology, 2002