Responses of Bilevel Ventilators to Unintentional Leak: A Bench Study
Open Access
- 30 November 2022
- journal article
- research article
- Published by MDPI AG in Healthcare
- Vol. 10 (12) , 2416
Abstract
Background: The impact of leaks has mainly been assessed in bench models using continuous leak patterns which did not reflect real-life leakage. We aimed to assess the impact of the pattern and intensity of unintentional leakage (UL) using several respiratory models. Methods: An active artificial lung (ASL 5000) was connected to three bilevel-ventilators set in pressure mode; the experiments were carried out with three lung mechanics (COPD, OHS and NMD) with and without upper airway obstruction. Triggering delay, work of breathing, pressure rise time, inspiratory pressure, tidal volume, cycling delay and the asynchrony index were measured at 0, 6, 24 and 36 L/min of UL. We generated continuous and inspiratory UL. Results: Compared to 0 L/min of UL, triggering delays were significantly higher with 36 L/min of UL (+27 ms) and pressure rise times were longer (+71 ms). Cycling delays increased from −4 [−250–169] ms to 150 [−173–207] ms at, respectively 0 L/min and 36 L/min of UL and work of breathing increased from 0.15 [0.12–0.29] J/L to 0.19 [0.16–0.36] J/L. Inspiratory leakage pattern significantly increased triggering delays (+35 ms) and cycling delays (+263 ms) but decreased delivered pressure (−0.94 cmH2O) compared to continuous leakage pattern. Simulated upper airway obstruction significantly increased triggering delay (+199 ms), cycling delays (+371 ms), and decreased tidal volume (−407 mL) and pressure rise times (−56 ms). Conclusions: The pattern of leakage impacted more the device performances than the magnitude of the leakage per se. Flow limitation negatively reduced all ventilator performances.Keywords
Funding Information
- Air Liquide Medical Systems
This publication has 35 references indexed in Scilit:
- Effect of dynamic random leaks on the monitoring accuracy of home mechanical ventilators: a bench studyBMC Pulmonary Medicine, 2013
- An Official American Thoracic Society Statement: Continuous Positive Airway Pressure Adherence Tracking Systems. The Optimal Monitoring Strategies and Outcome Measures in AdultsAmerican Journal of Respiratory and Critical Care Medicine, 2013
- Leaks can dramatically decrease FiO2 on home ventilators: a bench studyBMC Research Notes, 2013
- Effect of Leak and Breathing Pattern on the Accuracy of Tidal Volume Estimation by Commercial Home Ventilators: A Bench StudyRespiratory Care, 2013
- Asynchronies and sleep disruption in neuromuscular patients under home noninvasive ventilationRespiratory Medicine, 2012
- Air Leak is Associated with Poor Adherence to Autopap TherapySLEEP, 2011
- A Bench Study of the Effects of Leak on Ventilator Performance During Noninvasive VentilationRespiratory Care, 2011
- Ventilator modes and settings during non-invasive ventilation: effects on respiratory events and implications for their identificationThorax, 2010
- Evaluating noninvasive ventilation using a monitoring system coupled to a ventilator: a bench-to-bedside studyEuropean Respiratory Journal, 2009
- Effect of sleep on patient/ventilator asynchrony in patients undergoing chronic non-invasive mechanical ventilationRespiratory Medicine, 2007