An optoelectric plantar "shear" sensing transducer: design, validation, and preliminary subject tests
Authors:
- G.F. HarrisGHG.F. Harris
- J.J. WertschJWJ.J. Wertsch
- Hongsheng ZhuHZHongsheng Zhu
- 1 December 1996
- journal article
- review article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Rehabilitation Engineering
- Vol. 4 (4) , 310-319
Abstract
A prototype miniature plantar shear sensing transducer was developed, characterized, and tested in this study. Electro-optical components were chosen for the design because of the fast response time, low cost, small size, low power requirements, and adaptability to this application. The optoelectric circuit employed a 660 nm wavelength light source and photodiode solar cell. Signal amplification and sensitivity were adjusted to provide an output voltage proportional to light power. The sensor shell was designed to encapsulate the electro-optical sensing components while providing mechanical resistance to shear through a spring mechanism. A naval bronze was chosen for the shell due to its strength and nonreflective characteristics (alloy of copper and tin). Static and dynamic characteristics of the shear sensor were determined through a series of calibration tests. Mechanical crosstalk sensitivity ranged from 14.34 to 30.51 mV/N. This represented 1% full-scale/Newton sensitivity. Nonlinearity averaged 5.6% in the forward direction and 7.6% in the reverse direction. Overall sensor output hysteresis was 1.1 +/- 3.1% while the natural frequency of the sensor to an input shear transient was approximately 5 Hz. Temperature sensitivity was -7.0 mV/degree C or 3.5% full-scale/degree C. Testing of five adult subjects revealed peak anterior-posterior shear ranging from 6.7 kPa (posterior heel) to 51.4 kPa (great toe) and medial-lateral shear ranging from 5.4 kPa (great toe) to 43.5 kPa (first metatarsal head). Stress-time integral values ranged from 0.78 kPa-sec (posterior shear at the posterior heel) to 37.3 kPa-sec (medial shear at the posterior heel). Contact durations ranged from 0.28 sec (posterior shear at the posterior heel) to 1.25 sec (medial shear at the posterior heel). Further application of the sensor for plantar shear characterization in able-bodied subjects and those with pathology is suggested.Keywords
This publication has 13 references indexed in Scilit:
- Method for in-shoe shear stress measurementJournal of Biomedical Engineering, 1992
- The development of the low profile Liverpool shear transducerClinical Physics and Physiological Measurement, 1992
- A portable insole plantar pressure measurement systemJournal of Rehabilitation Research and Development, 1992
- A microprocessor-based data-acquisition system for measuring plantar pressures from ambulatory subjectsIEEE Transactions on Biomedical Engineering, 1991
- Study of the relative timing of shear forces on the sole of the forefoot during walkingJournal of Biomedical Engineering, 1991
- Diabetic foot lesionsJournal of Wound, Ostomy & Continence Nursing, 1990
- A miniature electrooptical force transducerIEEE Transactions on Biomedical Engineering, 1988
- The frequency content of gaitJournal of Biomechanics, 1985
- Forces under the footJournal of Biomedical Engineering, 1983
- Method of measuring 'shearing' forces on the sole of the footClinical Physics and Physiological Measurement, 1980