Response evaluation of two commercial thermoluminescence dosimeters (TLDs) against different parameters
Open Access
- 1 August 2023
- journal article
- research article
- Published by Oxford University Press (OUP) in BJR|Open
- Vol. 5 (1) , 20220035
Abstract
Objectives: It is essential to study the dosimetric performance and reliability of personal dosimeters. This study examines and compares the responses of two commercial thermoluminescence dosimeters (TLDs), the TLD-100 and the MTS-N. Methods: We compared the two TLDs to various parameters such as energy dependence, linearity, homogeneity, reproducibility, light sensitivity (zero point), angular dependence, and temperature effects using the IEC 61066 standard. Results: The results acquired showed that both TLD materials show linear behavior as indicated by the quality of the fit. In addition, the angular dependence results for both detectors show that all dose responses are within the range of acceptable values. However, the TLD-100 outperformed the MTS-N in terms of light sensitivity reproducibility for all detectors together, while the MTS-N outperforms the TLD-100 for each detector independently and that showed TLD-100 has more stability than MTS-N. The MTS-N shows better batch homogeneity (10.84%) than TLD-100 (13.65%). The effect of temperature in signal loss was clearer at higher temperature 65°C and it was however below ±30%. Conclusions: The overall results for dosimetric properties determined in terms of dose equivalents for all combinations of detectors are satisfactory. The MTS-N cards have better results in the energy dependence, angular dependency, batch homogeneity and less signal fading, whereas the TLD-100 cards are less sensitive to light and more reproducible. Advances in knowledge: Although previous studies showed several types of comparisons between TLDs, they have used limited parameters and different data analysis. This study has dealt with more comprehensive characterization methods and examinations combining TLD-100 and MTS-N cards.Keywords
This publication has 18 references indexed in Scilit:
- Analysis of dosimetric peaks of MgB4O7:Dy (40% Teflon) versus LiF:Mg,Ti TL detectorsNukleonika, 2016
- Response of TLD‐100 in mixed fields of photons and electronsMedical Physics, 2013
- COMPARISON OF TWO DIFFERENT TYPES OF LIF:MG,CU,P THERMOLUMINESCENT DOSIMETERS FOR DETECTION OF BETA RAYS (BETA-TLDS) FROM 90SR/90Y, 85KR AND 147PM SOURCESHealth Physics, 2011
- Response of TL materials to diagnostic radiology X radiation beamsApplied Radiation and Isotopes, 2010
- Experience of type testing Harshaw advanced TLD system to new IEC-61066 standardRadiation Measurements, 2008
- Comparison of LiF (TLD-100 and TLD-100H) detectors for extremity monitoringRadiation Measurements, 2008
- RADIONUCLIDE AND RADIATION PROTECTION DATA HANDBOOK 2002Radiation Protection Dosimetry, 2002
- Performance and Test Results of Harshaw Pelletised LiF:Mg,Ti TLD MaterialRadiation Protection Dosimetry, 1999
- Development of TL dosimeters based on MTS-N (LiF:Mg, Ti) detectors for in vivo dosimetry in a Co-60 beamReports of Practical Oncology and Radiotherapy, 1998
- A comparison of the effectiveness of thermoluminescent crystals LiF:Mg,Ti, and LiF:Mg,Cu,P for clinical dosimetryMedical Physics, 1997