Closed loop control of microscopic particles incorporating steady streaming and visual feedback
Authors:
- 9 March 2018
- journal article
- research article
- Published by Springer Nature in Biomedical Microdevices
- Vol. 20 (2) , 28
Abstract
Automatic manipulation of microscopic particles is very important in biology, especially in new lab-on-chip systems for automatic testing and DNA manipulation. We suggest a particle manipulation system (PMS) based on vibrating piezoelectric beams creating steady streaming flow in a viscous liquid. The flow is nearly unidirectional and it is used to control the position and velocity of the particles in the workspace of the PMS. The particles position in the PMS are controlled by visual feedback. This study presents the manipulation method, the system’s model describing its behavior and characterizes experimentally its performance. The PMS is capable moving a 2-200 μm particle in a workspace of 8x8 mm2 with an absolute accuracy of 0.2 μm. The characteristic velocity in 500 cP Si oil, is 20 μm/s using an actuation voltage amplitude of 5 V and can reach 250 μm/s using 15 V respectively. We can also move a constellation of several particles in various sizes without changing the distance between them. The accuracy of the manipulation can be increased by enhancing the amplification of the microscope on the expanse of a smaller workspace field of view.Keywords
This publication has 44 references indexed in Scilit:
- Bio-inspired magnetic swimming microrobots for biomedical applicationsNanoscale, 2012
- A non-contact method for spatially localized sedimentation of particles from liquid suspensions using Marangoni forcesJournal of Micromechanics and Microengineering, 2011
- Microrobots for Minimally Invasive MedicineAnnual Review of Biomedical Engineering, 2010
- High-precision tracking of sperm swimming fine structure provides strong test of resistive force theoryJournal of Experimental Biology, 2010
- Motion Control at the NanoscaleSmall, 2010
- Laser-Induced Thermocapillary Convection for Mesoscale ManipulationInternational Journal of Optomechatronics, 2009
- Flagellated Magnetotactic Bacteria as Controlled MRI-trackable Propulsion and Steering Systems for Medical Nanorobots Operating in the Human MicrovasculatureThe International Journal of Robotics Research, 2009
- Manipulation of biological samples using micro and nano techniquesIntegrative Biology, 2009
- Hydrodynamic Tweezers: 1. Noncontact Trapping of Single Cells Using Steady Streaming MicroeddiesAnalytical Chemistry, 2006
- Microscopic artificial swimmersNature, 2005