Evaluation of a photoplethysmography heart rate sensor: reliability and accuracy of the Fitbit Charge 4 in ecological conditions. (Preprint)
Authors:
- Maxime CeugniezMCMaxime Ceugniez
- Hervé DevanneHDHervé Devanne
- Eric HermandEric Hermand
Preprint
- 25 November 2023
- preprint
- Published by JMIR Publications Inc. in JMIR Preprints
Abstract
BACKGROUND The use of wrist-worn photoplethysmography (PPG) heart rate (HR) sensors has been quickly increasing over the recent years. Wearables could be useful for researchers and individuals for their simplicity to set up and use. However, only a limited number of studies investigated the reliability and accuracy of these devices in non-laboratory controlled conditions. OBJECTIVE The purpose of this study was (1) to assess the accuracy and reliability of the PPG based HR sensor of the Fitbit Charge 4 (FC4) in ecological condition; (2) to quantify the potential variability caused by the nature of the activity. METHODS HR was simultaneously collected during various sport sessions with the FC4 and a Polar H10, as criterion. Skin tone was assessed with Fitzpatrick Scale. Intraclass correlation coefficients (ICC), Lin’s concordance correlation coefficient (CCC) and Bland-Altman test determined reliability and accuracy of the sensor. RESULTS HR data of 26 students aged 21.1±5.8 years on averaged were recorded during 77.47 hours of physical activity. ICCs (and CCCs) for running and orienteering running were 0.900 (0.999) and 0.801 (0.932) respectively, whereas for badminton, tennis, bike and soccer, they were 0.365 (0.778), 0.421 (0.884), 0.658 (0.971) and 0.487 (0.809), respectively. Highest biases were found for badminton and soccer, whereas smallest was found for running. CONCLUSIONS Researchers and individuals should not use the FC4 for measuring HR because of its LOA width and sensitivity to motion artifacts.Keywords
This publication has 32 references indexed in Scilit:
- Advances in photoplethysmography: beyond arterial oxygen saturationCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 2015
- MICROST: A mixed approach for heart rate monitoring during intensive physical exercise using wrist-type PPG Signals2015
- PhotoplethysmographyBest Practice & Research Clinical Anaesthesiology, 2014
- Comparison between red, green and blue light reflection photoplethysmography for heart rate monitoring during motion2013
- Influence of skin type and wavelength on light wave reflectanceJournal of Clinical Monitoring and Computing, 2013
- Photoplethysmography and its application in clinical physiological measurementPhysiological Measurement, 2007
- Influence of Contact Forces on Wrist Photo plethysmography – Prestudy for a Wearable Patient Monitor / Einfluß von Kontaktkräften auf die photoplethysmografische Pulsmessung am Handgelenk – Vorstudie zu einem tragbaren PatientenmonitorBiomedical Engineering / Biomedizinische Technik, 2004
- The Validity and Practicality of Sun-Reactive Skin Types I Through VIArchives of Dermatology, 1988
- Pulsed multifrequency photoplethysmographMedical & Biological Engineering & Computing, 1984
- Photoelectric Plethysmography of the Fingers and Toes in ManExperimental Biology and Medicine, 1937