Kinetics of arterial carbon dioxide during veno-venous extracorporeal membrane oxygenation support in an apnoeic porcine model
Authors:
- Pedro Vitale MendesPMPedro Vitale Mendes
- Marcelo ParkMPMarcelo Park
- Alexandre Toledo MacielAMAlexandre Toledo Maciel
- Débora Prudêncio e SilvaDSDébora Prudêncio e Silva
- Natalia FriedrichNFNatalia Friedrich
- Edzangela Vasconcelos Santos BarbosaEBEdzangela Vasconcelos Santos Barbosa
- Adriana Sayuri HirotaAHAdriana Sayuri Hirota
- Guilherme Pinto Paula SchettinoGSGuilherme Pinto Paula Schettino
- Luciano Cesar Pontes AzevedoLALuciano Cesar Pontes Azevedo
- Eduardo Leite Vieira CostaECEduardo Leite Vieira Costa
Open Access
- 6 January 2016
- journal article
- research article
- Published by Springer Nature in Intensive Care Medicine Experimental
- Vol. 4 (1) , 1-11
Abstract
Extracorporeal membrane oxygenation (ECMO) is a technique widely used worldwide to improve gas exchange. Changes in ECMO settings affect both oxygen and carbon dioxide. The impact on oxygenation can be followed closely by continuous pulse oximeter. Conversely, carbon dioxide equilibrates much slower and is not usually monitored directly. We investigated the time to stabilization of arterial carbon dioxide partial pressure (PaCO2) following step changes in ECMO settings in 5 apnoeic porcine models under veno-venous ECMO support with polymethylpentene membranes. We collected sequential arterial blood gases at a pre-specified interval of 50 min using a sequence of standardized blood and sweep gas flow combinations. Following the changes in ECMO parameters, the kinetics of carbon dioxide was dependent on sweep gas and ECMO blood flow. With a blood flow of 1500 mL/min, PaCO2 takes longer than 50 min to equilibrate following the changes in sweep gas flow. Furthermore, the sweep gas flow from 3.0 to 10.0 L/min did not significantly affect PaCO2. However, with a blood flow of 3500 mL/min, 50 min was enough for PaCO2 to reach the equilibrium and every increment of sweep gas flow (up to 10.0 L/min) resulted in additional reductions of PaCO2. Fifty minutes was enough to reach the equilibrium of PaCO2 after ECMO initiation or after changes in blood and sweep gas flow with an ECMO blood flow of 3500 ml/min. Longer periods may be necessary with lower ECMO blood flows.Keywords
This publication has 21 references indexed in Scilit:
- Extracorporeal Membrane Oxygenation for Pandemic Influenza A(H1N1)–induced Acute Respiratory Distress SyndromeAmerican Journal of Respiratory and Critical Care Medicine, 2013
- Referral to an Extracorporeal Membrane Oxygenation Center and Mortality Among Patients With Severe 2009 Influenza A(H1N1)JAMA, 2011
- Troubleshooting adult ECMO.2011
- Extracorporeal Membrane Oxygenation for Treating Severe Cardiac and Respiratory Failure in Adults: Part 2—Technical ConsiderationsJournal of Cardiothoracic and Vascular Anesthesia, 2010
- Extracorporeal Membrane Oxygenation for 2009 Influenza A(H1N1) Acute Respiratory Distress SyndromeJAMA, 2009
- High survival in adult patients with acute respiratory distress syndrome treated by extracorporeal membrane oxygenation, minimal sedation, and pressure supported ventilationIntensive Care Medicine, 2000
- The distribution and physiological significance of carbonic anhydrase in vertebrate gas exchange organsRespiration Physiology, 2000
- Calculation of whole blood CO2 contentJournal of Applied Physiology, 1988
- Oxygen and carbon dioxide gas stores of the body.Physiological Reviews, 1970
- The state of carbon dioxide in bloodThe Journal of Physiology, 1933