An in situ gas chromatograph with automatic detector switching between PTR- and EI-TOF-MS: isomer-resolved measurements of indoor air
Open Access
- 7 January 2021
- journal article
- research article
- Published by Copernicus GmbH in Atmospheric Measurement Techniques
- Vol. 14 (1) , 133-152
Abstract
We have developed a field-deployable gas chromatograph (GC) with thermal desorption preconcentration (TDPC), which is demonstrated here with automatic detector switching between two high-resolution time-of-flight mass spectrometers (TOF-MSs) for in situ measurements of volatile organic compounds (VOCs). This system provides many analytical advances, including acquisition of fast time–response data in tandem with molecular speciation and two types of mass spectral information for each resolved GC peak: molecular ion identification from Vocus proton transfer reaction (PTR) TOF-MS and fragmentation pattern from electron ionization (EI) TOF-MS detection. This system was deployed during the 2018 ATHLETIC campaign at the University of Colorado Dal Ward Athletic Center in Boulder, Colorado, where it was used to characterize VOC emissions in the indoor environment. The addition of the TDPC-GC increased the Vocus sensitivity by a factor of 50 due to preconcentration over a 6 min GC sample time versus direct air sampling with the Vocus, which was operated with a time resolution of 1 Hz. The GC-TOF methods demonstrated average limits of detection of 1.6 ppt across a range of monoterpenes and aromatics. Here, we describe the method to use the two-detector system to conclusively identify a range of VOCs including hydrocarbons, oxygenates, and halocarbons, along with detailed results including the quantification of anthropogenic monoterpenes, where limonene accounted for 47 %–80 % of the indoor monoterpene composition. We also report the detection of dimethylsilanediol (DMSD), an organosiloxane degradation product, which was observed with dynamic temporal behavior distinct from volatile organosiloxanes (e.g., decamethylcyclopentasiloxane, D5 siloxane). Our results suggest DMSD is produced from humidity-dependent heterogeneous reactions occurring on surfaces in the indoor environment, rather than formed through gas-phase oxidation of volatile siloxanes.Keywords
All Related Versions
Funding Information
- Alfred P. Sloan Foundation (G-2016-7173, G-2019-12444)
This publication has 61 references indexed in Scilit:
- Long-Term Trends in Motor Vehicle Emissions in U.S. Urban AreasEnvironmental Science & Technology, 2013
- The Model of Emissions of Gases and Aerosols from Nature version 2.1 (MEGAN2.1): an extended and updated framework for modeling biogenic emissionsGeoscientific Model Development, 2012
- Temperature-Dependent Vapor Pressure of Selected Cyclic and Linear Polydimethylsiloxane OligomersJournal of Chemical & Engineering Data, 2010
- Analyses of volatile organic compounds from human skinBritish Journal of Dermatology, 2008
- Measurements of volatile organic compounds in the earth's atmosphere using proton‐transfer‐reaction mass spectrometryMass Spectrometry Reviews, 2006
- Validation of proton transfer reaction‐mass spectrometry (PTR‐MS) measurements of gas‐phase organic compounds in the atmosphere during the New England Air Quality Study (NEAQS) in 2002Journal of Geophysical Research: Atmospheres, 2003
- Degradation of Silicone Polymer at Different Soil MoisturesEnvironmental Science & Technology, 1998
- Ionization cross-sections and ionization efficiency curves from polarizability volumes and ionization potentialsInternational Journal of Mass Spectrometry and Ion Processes, 1997
- Study of the Degradation of Polydimethylsiloxanes on SoilEnvironmental Science & Technology, 1995
- Degradation of silicone polymers in soilEnvironmental Toxicology and Chemistry, 1994