Gain Calibration of Current-to-Voltage Converters
Authors:
- 19 November 2018
- journal article
- research article
- Published by National Institute of Standards and Technology (NIST) in Journal of Research of the National Institute of Standards and Technology
- Vol. 123, 1-15
Abstract
Current-to-voltage converters are used in many photometric and radiometric applications. The calibration of current-to-voltage converters at a few input currents is not always sufficient to understand the linearity and the bias of a device. Many devices have structure deviating from a linear response over the operating range of a gain setting. Measurement services that rely on these devices now have decreased uncertainties to a level that requires quantifying the uncertainties and understanding how they propagate. The National Institute of Standards and Technology has developed a system to calibrate the current-to-voltage conversion factor or “gain” and offset of these devices for direct current photocurrents. The equipment used for the calibration is described here, and the results and uncertainties are discussed.Keywords
Funding Information
- Physical Measurement Laboratory
This publication has 5 references indexed in Scilit:
- Development of an in situ calibration method for current-to-voltage converters for high-accuracy SI-traceable low dc current measurementsMetrologia, 2013
- Calibrations of current-to-voltage transimpedance amplifiers using electrical standards2010
- Traceability of photocurrent measurements to electrical standardsMAPAN, 2009
- NIST measurement service for DC standard resistors2003
- A 10-V Josephson voltage standardIEEE Transactions on Instrumentation and Measurement, 1989