Evaluation of the Versius Robotic System for Infant Surgery—A Study in Piglets of Less than 10 kg Body Weight
Authors:
- Thomas Franz KrebsThomas Franz Krebs
- Timo KayserTKTimo Kayser
- Ulf LorenzenULUlf Lorenzen
- Matthias GrünewaldMGMatthias Grünewald
- Marit KayserMKMarit Kayser
- Anna SaltnerAnna Saltner
- Lidya-Olgu DurmazLidya-Olgu Durmaz
- Lina Johanna ReeseLRLina Johanna Reese
- Ewan BrownleeEBEwan Brownlee
- Katja ReischigKRKatja Reischig
- Jonas BaastrupJBJonas Baastrup
- Andreas MeinzerAMAndreas Meinzer
Open Access
Abstract
Background: We were able to demonstrate the feasibility of a new robotic system (Versius, CMR Surgical, Cambridge, UK) for procedures in small inanimate cavities. The aim of this consecutive study was to test the Versius® system for its feasibility, performance, and safety of robotic abdominal and thoracic surgery in piglets simulating infants with a body weight lower than 10 kg. Methods: A total of 24 procedures (from explorative laparoscopy to thoracoscopic esophageal repair) were performed in 4 piglets with a mean age of 12 days and a mean body weight of 6.4 (7–7.5) kg. Additional urological procedures were performed after euthanasia of the piglet. The Versius® robotic system was used with 5 mm wristed instruments and a 10 mm 3D 0° or 30° camera. The setup consisted of the master console and three to four separate arms. The performance of the procedure, the size, position, and the distance between the ports, the external and internal collisions, and complications of the procedures were recorded and analyzed. Results: We were able to perform all surgical procedures as planned. We encountered neither surgical nor robot-associated complications in the live model. Whereas all abdominal procedures could be performed successfully under general anesthesia, one piglet was euthanized early before the thoracic interventions, likely due to pulmonary inflammatory response. Technical limitations were based on the size of the camera (10 mm) being too large and the minimal insertion depth of the instruments for calibration of the fulcrum point. Conclusions: Robotic surgery on newborns and infants appears technically feasible with the Versius® system. Software adjustments for fulcrum point calibration need to be implemented by the manufacturer as a result of our study. To further evaluate the Versius® system, prospective trials are needed, comparing it to open and laparoscopic surgery as well as to other robotic systems.Keywords
Funding Information
- CMR Surgical (CMR Surgical Ltd, 194 Science Park, Milton Road, Cambridge, UK)
This publication has 28 references indexed in Scilit:
- da Vinci robot-assisted keyhole neurosurgery: a cadaver study on feasibility and safetyNeurosurgical Review, 2014
- Advances in Minimally Invasive Surgery in Pediatric PatientsAdvances in Pediatrics, 2014
- Improving Bioscience Research Reporting: The ARRIVE Guidelines for Reporting Animal ResearchPLOS Biology, 2010
- Robotic Surgery in Small Children: Is There Room for This?Journal of Laparoendoscopic & Advanced Surgical Techniques, 2009
- No surgical innovation without evaluation: the IDEAL recommendationsThe Lancet, 2009
- Morgagni Hernia Repair in a Small Child Using da Vinci Robotic Instruments – A Case ReportEuropean Journal of Pediatric Surgery, 2008
- Psychomotor performance measured in a virtual environment correlates with technical skills in the operating roomSurgical Endoscopy, 2008
- Is Smaller Workspace a Limitation for Robot Performance in Laparoscopy?Journal of Urology, 2008
- Pediatric robotic surgery: A single-institutional review of the first 100 consecutive casesSurgical Endoscopy, 2007
- Testing technical skill via an innovative “bench station” examinationThe American Journal of Surgery, 1997