Combining C-arm CT with a new remote operated positioning and guidance system for guidance of minimally invasive spine interventions
Authors:
- 5 March 2014
- journal article
- Published by BMJ in Journal of NeuroInterventional Surgery
- Vol. 7 (4) , 303-308
Abstract
Objective To report our experience using C-arm cone beam CT (C-arm CBCT) combined with the new remote operated positioning and guidance system, iSYS1, for needle guidance during spinal interventions. Methods A C-arm CBCT with a flat panel angiography system was acquired (Artis Zeego; Siemens Healthcare Sector, Forchheim, Germany). Reconstruction of CT-like images and planning of the needle path were performed using a common workstation. The needle holder of iSYS1 acted as a guide during insertion of Kirschner (K) wires. 20 percutaneous K wires were placed in the pedicles at T2–T3, T7–T12, and L1–L2 in a cadaver specimen. Postprocedure C-arm CBCT scans were obtained to confirm the accuracy of the K wire placement. Results All K wire placements were successfully performed. Mean planning time with Syngo iGuide was 4:16 min, mean positioning time of iSYS1 was 3:35 min, and mean placement time of the K wires was 2:22 min. Mean total intervention time was 10:13 min per pedicle. A mean deviation of 0.35 mm between the planned path and the placed K wire with a mean path length of 6.73 cm was documented. Conclusions Our results demonstrate the potential of combining C-arm CBCT with iSYS1 for safe and accurate percutaneous placement of pedicle K wires in spinal interventions.Keywords
This publication has 23 references indexed in Scilit:
- The Accuracy of Navigation and 3D Image-Guided Placement for the Placement of Pedicle Screws in Congenital Spine DeformityJournal of Pediatric Orthopaedics, 2012
- Pedicle screw navigation: a systematic review and meta-analysis of perforation risk for computer-navigated versus freehand insertionJournal of Neurosurgery: Spine, 2012
- Computed Tomography–Guided Navigation of Thoracic Pedicle Screws for Adolescent Idiopathic Scoliosis Results in More Accurate Placement and Less Screw RemovalSpine, 2012
- Clinical Assessment of Percutaneous Lumbar Pedicle Screw Placement Using the O-Arm Multidimensional Surgical Imaging SystemNeurosurgery, 2012
- Accuracy and Safety in Pedicle Screw Placement in the Thoracic and Lumbar Spines : Comparison Study between Conventional C-Arm Fluoroscopy and Navigation Coupled with O-Arm® Guided MethodsJournal of Korean Neurosurgical Society, 2012
- Pedicle Screw Placement With O-arm and Stealth NavigationOrthopedics, 2012
- Accuracy of pedicle screw placement: a systematic review of prospective in vivo studies comparing free hand, fluoroscopy guidance and navigation techniquesEuropean Spine Journal, 2011
- C-Arm Cone Beam Computed Tomography Needle Path Overlay for Fluoroscopic Guided VertebroplastySpine, 2010
- Pedicle Screw Placement in the Thoracic SpineSpine, 2005
- Pedicle Screw Instrumentation of the Thoracic Spine in Idiopathic ScoliosisSpine, 1997