Abstract
The point-by point 3D scanning strategy adopted in Stimulated Emission Depletion Microscopy (STED) is time-consuming. The 3D scanning can be replaced with a 2D scanning in the non-diffracting Bessel-Bessel STED (BB-STED). In order to extract the excited emitters' axial information in BB-STED, we propose to encode axial information by using a detection optical path with single-helix PSF, and then predict the depths of the emitters with deep learning. Simulation demonstrated that, for dense emitters in a depth range of 4 μm, an axial precision of ∼35 nm can be achieved. Our method also works for experimental data, and an axial precision of ∼63 nm can be achieved.
Funding Information
  • National Natural Science Foundation of China (62475168, 11774242, 62175166, 61335001)
  • National Key Research and Development Program of China (2022YFF0712500)
  • Shenzhen Science and Technology Planning (JCYJ20210324094200001, JCYJ20200109105411133)
  • Shenzhen Key Laboratory of Photonics and Biophotonics (ZDSYS20210623092006020)