Analysis of bacterial contamination and the effectiveness of UV light-based reprocessing of everyday medical devices
Open Access
- 9 November 2022
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 17 (11) , e0268863
Abstract
Background: The reprocessing of daily used medical devices is often inadequate, making them a potential source of infection. In addition, there are usually no consistent and technically standardized procedures available for this purpose. Hence, the aim of this study is to analyze the bacterial contamination and the effectiveness of Ultraviolet light-based (UV light-based) reprocessing of daily used medical devices. Material and methods: Six different everyday medical devices (20 each; stethoscopes, tourniquets, bandage scissors, reflex hammers, tuning forks, and nystagmus glasses) were tested for bacterial contamination. All medical devices were then exposed to UV-C light for 25 seconds. Medical devices with a smooth surface were pre-cleaned with a water-based wipe. Contact samples were taken before and after reprocessing. Results: Immediately after clinical use, 104 of 120 contact samples showed an average bacterial contamination of 44.8±64.3 colony forming units (CFU) (0–300 CFU), also including potentially pathogenic bacteria. Two further culture media were completely overgrown with potentially pathogenic bacteria. The stethoscopes were found to have the highest average contamination of 90±91.6 CFU. After reprocessing, 118 of 120 samples were sterile, resulting in an average residual contamination of 0.02±0.1 CFU in two samples, whereby only bacteria of the ordinary skin flora were found. Conclusion: The present study shows the potentially clinically relevant bacterial contamination of everyday used medical devices. The reprocessing method tested here using UV light appears to be a suitable method for disinfection, especially for objects that up to now have been difficult to disinfect or cannot be disinfected in a standardized manner.Keywords
Funding Information
- UV Smart Technologies B.V. (N/A)
- UV Smart Technologies B.V. (N/A)
This publication has 35 references indexed in Scilit:
- 2007 Guideline for Isolation Precautions: Preventing Transmission of Infectious Agents in Health Care SettingsAmerican Journal of Infection Control, 2007
- Bacterial contamination of stethoscopes with antimicrobial diaphragm coversAmerican Journal of Infection Control, 2007
- A strategy to reduce MRSA colonization of stethoscopesJournal of Hospital Infection, 2006
- A cluster of nosocomial Klebsiella pneumoniae bloodstream infections in a neonatal intensive care department: identification of transmission and interventionAmerican Journal of Infection Control, 2003
- Scissors: A Potential Source of Nosocomial InfectionInfection Control & Hospital Epidemiology, 2002
- Stethoscope: a friend or an enemy?Sao Paulo Medical Journal, 2002
- Lapses in measures recommended for preventing hospital-acquired infectionJournal of Hospital Infection, 2001
- The stethoscope. A potential source of nosocomial infection?Archives of internal medicine (1960), 1997
- Ultraviolet inactivation of selected bacteria and viruses with photoreactivation of the bacteriaWater Research, 1987
- Anticipatics, handwashing, and handwashing facililtiesAmerican Journal of Infection Control, 1983