Pseudotumor with Superimposed Periprosthetic Infection Following Metal-on-Metal Total Hip Arthroplasty
- 7 July 2010
- journal article
- case report
- Published by Wolters Kluwer Health in Journal of Bone and Joint Surgery
- Vol. 92 (7) , 1666-1669
Abstract
Second-generation metal-on-metal bearing surfaces made of modern cobalt-chromium-molybdenum alloys are widely used for total hip arthroplasty in the United States1. While these bearing surfaces offer considerable advantages over conventional metal-on-polyethylene articulations, metal hypersensitivity reactions to these implants are an important, although uncommon, cause of failure2. This unique mode of failure, which has been reported with first-generation implants2,3 and subsequently in association with second-generation metal-on-metal bearing surfaces, appears to be caused by an immunologic delayed hypersensitivity response to metal particles4-6. This local tissue reaction, described as an aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL), was characterized by Willert et al.7 as having histologic features that include diffuse perivascular infiltrates of T and B lymphocytes and plasma cells, high endothelial venules, massive fibrin exudation, accumulation of macrophages with droplike inclusions, and infiltrates of eosinophilic granulocytes and necrosis. Histologically, high endothelial venules are characterized as vascular and/or lymphatic vessels with reactive endothelium, often seen in association with an inflammatory response. Willert et al. suggested that the possibility of such a reaction be considered when, following primary implantation of metal-on-metal implants, a patient reports the recurrence of preoperative pain and there is a marked joint effusion or the development of osteolysis in the absence of a suspicion of deep periprosthetic infection. A subsequent report, however, described this hypersensitivity reaction mimicking infection in a patient with elevated levels of inflammatory biomarkers who was later found to have no deep infection on revision8. We describe a patient with a failure of a metal-on-metal total hip replacement consistent with a metal hypersensitivity reaction who presented with a pseudotumor and massive lower-extremity edema secondary to extrinsic femoral vein compression and subsequently developed a superimposed deep periprosthetic infection prior to revision. To our knowledge, this has not …Keywords
This publication has 15 references indexed in Scilit:
- The Epidemiology of Bearing Surface Usage in Total Hip Arthroplasty in the United StatesJournal of Bone and Joint Surgery, 2009
- Failure of Metal-on-Metal Total Hip Arthroplasty Mimicking Hip InfectionJournal of Bone and Joint Surgery, 2009
- Cystic Lesion of the Groin due to Metallosis: A Rare Long-Term Complication of Metal-on-Metal Total Hip ArthroplastyThe Journal of Arthroplasty, 2007
- Survivorship and Retrieval Analysis of Sikomet Metal-on-Metal Total Hip Replacements at a Mean of Seven YearsJournal of Bone and Joint Surgery, 2006
- Metallosis After Contemporary Metal-on-Metal Total Hip ArthroplastyJournal of Bone and Joint Surgery, 2006
- Early Osteolysis Following Second-Generation Metal-on-Metal Hip ReplacementJournal of Bone and Joint Surgery, 2005
- Metal-on-Metal Bearings and Hypersensitivity in Patients with Artificial Hip JointsJournal of Bone and Joint Surgery, 2005
- Cup inclination and serum concentration of cobalt and chromium after metal-on-metal total hip arthroplastyThe Journal of Arthroplasty, 2004
- Characteristics of degeneration in an unstable knee with a coronal surface step-offThe Journal of Bone and Joint Surgery. British volume, 2001
- Incidence of metal sensitivity in patients with total joint replacements.BMJ, 1975