Abstract
A spatially weighted filter applied to both the measurement and the Jacobian is proposed for high-density diffuse correlation tomography (DCT) to remove unwanted extracerebral interferences and artefacts along with better depth localization in the reconstructed blood flow images. High-density DCT is implemented by appropriate modification of recently introduced Multi-speckle Diffuse Correlation Spectroscopy (M-DCS) system. Additionally, we have used autocorrelation measurements at multiple delay-times in an iterative manner to improve the reconstruction results. The proposed scheme has been validated by simulations, phantom experiments and in-vivo human experiments.
Funding Information
  • Wadhwani Research Centre for Bioengineering (WRCB), Indian Institute of Technology Bombay
  • Department of Biotechnology, Ministry of Science and Technology, India (Ramalingaswamy Fellowship-2016)
  • Department of Science and Technology, Ministry of Science and Technology, India (SERB–Early career research award)
  • Indian Institute of Technology Bombay (SEED GRANT)

This publication has 49 references indexed in Scilit: