Abstract
Purpose: To measure the endotoxin concentration (EC) of 25 commercially available, hyaluronic acid- and hydroxypropylmethylcellulose-based (HPMC) ophthalmic viscosurgical devices (OVDs). Methods: The in vitro Limulus amebocyte lysate (LAL) assay, which indicates the presence of endotoxins originating from gram-negative bacteria, was used to determine the EC. The procedure was performed according to the European Pharmacopoeia/USP. EC including duplicate determinations, negative controls, dilution series with control standard endotoxin, dilution series with sample extract and positive sample control. Results: 16 OVDs (Amvisc®, Amvisc® Plus, Biolon®, Coatel®, Healon®, Healon® GV, Healon®5, HPMC Ophtal® L, Microvisc®, Microvisc® Plus, Ocucoat®, Provisc®, Rayvisc®, Viscoat®, Visco Shield® 2%, Visko® 1.4%) had an EC under 1.2 endotoxin units/mL, five (Adatocel®, HPMC Ophtal® H, LA Gel®, Viscorneal®, Viscorneal® Plus) had an EC ≥ 1.2 and ≤ 24 EU/ml, and four (Biocorneal®, Dispasan® also named Ophthalin, Dispasan® Plus, Visko® 1%) had an EC of > 24 EU/ml. Discussion: To avoid viscoelastic-related inflammatory or immunological reactions, the use of pure OVDs is recommended, especially for surgical procedures with an inherent possibility of leaving viscoelastic remnants in the eye (e.g., cataract surgery, viscocanalostomy or penetrating keratoplasty).Keywords
This publication has 10 references indexed in Scilit:
- Silicone Oil Bubbles in Ophthalmic Viscosurgical DevicesJournal of Cataract & Refractive Surgery, 2002
- Healon5 viscoadaptive formulation: Comparison to Healon and Healon GVJournal of Cataract & Refractive Surgery, 2001
- Healon5 versus Viscoat during cataract surgery: intraocular pressure, laser flare and corneal changesGraefe's Archive for Clinical and Experimental Ophthalmology, 2000
- Viscoelastics in Ophthalmic Surgery2000
- Viscoelastic PurityJournal of Cataract & Refractive Surgery, 1998
- Epidemic Bacillus Endophthalmitis after Cataract Surgery IOphthalmology, 1997
- Primary Structure of Anti-Lipopolysaccharide Factor from American Horseshoe Crab, Limulus polyphemus1The Journal of Biochemistry, 1987
- Corneal Subepithelial Deposits After Use of Sodium ChondroitinAmerican Journal of Ophthalmology, 1986
- Lipopolysaccharide‐sensitive serine‐protease zymogen (factor C) found in Limulus hemocytesEuropean Journal of Biochemistry, 1986
- A new (1 → 3)‐β‐D‐glucan‐mediated coagulation pathway found in limulus amebocytesFEBS Letters, 1981