Modular Dynamic Torso Phantom Featuring Respiratory and Cardiac Motion for Magnetic Resonance Imaging
Authors:
- Jean-Lynce GnanagoJGJean-Lynce Gnanago
- Ernesto Gomez TammETErnesto Gomez Tamm
- Andreas HodulAHAndreas Hodul
- Quang NguyenQNQuang Nguyen
- Markus OrtnerMOMarkus Ortner
- Zacharias ChalampalakisZCZacharias Chalampalakis
- Onisim SoancaOSOnisim Soanca
- Michael WoletzMWMichael Woletz
- Lena NohavaLNLena Nohava
- Karyna IsaievaKIKaryna Isaieva
- Elmar LaistlerELElmar Laistler
- Ivo RauschIRIvo Rausch
- 14 January 2025
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 74 (00189456) , 1-9
Abstract
The design, fabrication, assembly and testing of a modular dynamic torso phantom for magnetic resonance imaging (MRI) is presented. The phantom is composed of fully MRI-compatible materials and consists of a saline-filled acrylic shell, with dimensions comparable to a human torso, and three distinct interior motion modules. Two linear piezo stages are used to mimic respiratory motion of the chest-wall (≤ 10 mm; cycle time ≥ 2 s) and the abdominal organs (≤ 45 mm; cycle time ≥ 2 s). A pneumatic system is used to move a heart mock-up to represent cardiac motion (≤ 19 mm, rate ≤ 3 Hz). The motion capabilities and MRI compatibility of the phantom are tested in 3 T MRI experiments. The closed-loop control of the phantom’s breathing modules allows for accurate positioning and timing with absolute differences between input and output values ≤ 0.61 mm and ≤ 0.06 s. The purpose of this phantom is to serve as a scientific instrument for the future development and evaluation of motion tracking and artifact compensation techniques.Keywords
Funding Information
- Austrian Science Fund (P37189 OPTIMAL, P35305 AquaLactEMISM)
- Focus Grant T4MR from the Center for Medical Physics and Biomedical Engineering (CMPBME), Medical University of Vienna
This publication has 45 references indexed in Scilit:
- Motion modeling from 4D MR images of liver simulating phantomJournal of Applied Clinical Medical Physics, 2022
- MRIgRT dynamic lung motion thorax anthropomorphic QA phantom: Design, development, reproducibility, and feasibility studyMedical Physics, 2019
- Motion artifacts in MRI: A complex problem with many partial solutionsJournal of Magnetic Resonance Imaging, 2015
- Dynamic phantom with heart, lung, and blood motion for initial validation of MRI techniquesJournal of Magnetic Resonance Imaging, 2011
- Evaluation of a rapid, multiphase MRE sequence in a heart‐simulating phantomMagnetic Resonance in Medicine, 2009
- MR elastography as a method for the assessment of myocardial stiffness: Comparison with an established pressure–volume model in a left ventricular model of the heartMagnetic Resonance in Medicine, 2009
- Generalized Reconstruction by Inversion of Coupled Systems (GRICS) applied to free‐breathing MRIMagnetic Resonance in Medicine, 2008
- A Simple and Realistic Tissue-Equivalent Breast Phantom for MRIJournal of Magnetic Resonance Imaging, 1999
- MRI geometric distortion: A simple approach to correcting the effects of non-linear gradient fieldsJournal of Magnetic Resonance Imaging, 1999
- Physiologic motion phantom for MRI applicationsJournal of Magnetic Resonance Imaging, 1996