Abstract
Wireless power transfer is a widely-adopted approach to recharge batteries in implantable devices. For fast charging and reducing the size of the implanted receiver coil, a wireless power receiver needs to support high input power, P IN , and operate at a high resonance frequency, f res , while providing high power efficiency to avoid tissue overheating. These requirements however have various design tradeoffs. The prior resonant voltage-mode receiver was reported to operate at 13.56MHz for direct battery charging, but it adopted six thick-oxide power FETs and an extra 4.7μH inductor, resulting in large conduction loss and degraded power efficiency [1]. Although the resonant current-mode receiver, RCM-RX, shown in Fig. 1, has a simpler structure with only two power FETs M P and M N , it cannot store large P IN due to the concern of the allowable voltage stress on LV power transistors [2], [3]. While a higher P IN is supported in [4], the RCM-RX uses a total of 9 power FETs for an additional SC converter, impairing the power efficiency and operation frequency. The RCM-RX in [5] can operate at 13.56MHz, but its maximum P IN is limited to a relatively low level of 17.5mW due to its peak-voltage timing detection scheme, PVTD. The PVTD would trigger the charging phase control in the N R th cycle and is strongly dependent on the controller delay, t delay . As shown in Fig. 1, when t delay becomes more dominant at a higher frequency and is prolonged due to the use of large-size power FETs for supporting higher P IN, t ch would be significantly shortened, degrading the power efficiency of the RX.