Quantification of Blood Velocity with 4D Digital Subtraction Angiography Using the Shifted Least-Squares Method
Open Access
- 13 September 2018
- journal article
- research article
- Published by American Society of Neuroradiology (ASNR) in American Journal of Neuroradiology
- Vol. 39 (10) , 1871-1877
Abstract
BACKGROUND AND PURPOSE: 4D-DSA provides time-resolved 3D-DSA volumes with high temporal and spatial resolutions. The purpose of this study is to investigate a shifted least squares method to estimate the blood velocity from the 4D DSA images. Quantitative validation was performed using a flow phantom with an ultrasonic flow probe as ground truth. Quantification of blood velocity in human internal carotid arteries was compared with measurements generated from 3D phase-contrast MR imaging. MATERIALS AND METHODS: The centerlines of selected vascular segments and the time concentration curves of each voxel along the centerlines were determined from the 4D-DSA dataset. The temporal shift required to achieve a minimum difference between any point and other points along the centerline of a segment was calculated. The temporal shift as a function of centerline point position was fit to a straight line to generate the velocity. The proposed shifted least-squares method was first validated using a flow phantom study. Blood velocities were also estimated in the 14 ICAs of human subjects who had both 4D-DSA and phase-contrast MR imaging studies. Linear regression and correlation analysis were performed on both the phantom study and clinical study, respectively. RESULTS: Mean velocities of the flow phantom calculated from 4D-DSA matched very well with ultrasonic flow probe measurements with 11% relative root mean square error. Mean blood velocities of ICAs calculated from 4D-DSA correlated well with phase-contrast MR imaging measurements with Pearson correlation coefficient r = 0.835. CONCLUSIONS: The availability of 4D-DSA provides the opportunity to use the shifted least-squares method to estimate velocity in vessels within a 3D volume.Keywords
Funding Information
- Siemens Healthineers
- Center for Scientific Review (R01HL116567-05)
This publication has 22 references indexed in Scilit:
- A DSA-Based Method Using Contrast-Motion Estimation for the Assessment of the Intra-Aneurysmal Flow Changes Induced by Flow-Diverter StentsAmerican Journal of Neuroradiology, 2012
- Comparison of Phase-Contrast MR Imaging and Endovascular Sonography for Intracranial Blood Flow Velocity MeasurementsAmerican Journal of Neuroradiology, 2012
- Velocity Measurements in the Middle Cerebral Arteries of Healthy Volunteers Using 3D Radial Phase-Contrast HYPRFlow: Comparison with Transcranial Doppler Sonography and 2D Phase-Contrast MR ImagingAmerican Journal of Neuroradiology, 2010
- Color-Coded Digital Subtraction Angiography: The End of a Monochromatic Era?American Journal of Neuroradiology, 2010
- Parametric Color Coding of Digital Subtraction AngiographyAmerican Journal of Neuroradiology, 2010
- Transcranial Doppler Ultrasonography in Cerebral Arteriovenous MalformationsStroke, 1995
- Building Skeleton Models via 3-D Medial Surface Axis Thinning AlgorithmsCVGIP: Graphical Models and Image Processing, 1994
- Determination of Instantaneous and Average Blood Flow Rates from Digital Angiograms of Vessel Phantoms Using Distance-Density CurvesInvestigative Radiology, 1991
- A new algorithm for deriving pulsatile blood flow waveforms tested using simulated dynamic angiographic dataNeuroradiology, 1989
- Arterial Blood Flow Measurement: Assessment of Velocity Estimation MethodsInvestigative Radiology, 1977