Characterization and correction of system delays and eddy currents for MR imaging with ultrashort echo‐time and time‐varying gradients
Open Access
- 7 April 2009
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 62 (2) , 532-537
Abstract
Reconstruction of high‐quality MR images requires precise knowledge of the dynamic gradient magnetic fields used to perform spatial encoding. System delays and eddy currents can perturb the gradient fields in both time and space and significantly degrade the image quality for acquisitions with an ultrashort echo time or with rapidly varying readout gradient waveforms. A technique for simultaneously characterizing and correcting the system delay and linear‐ and zero‐order eddy currents of an MR system is proposed. A single set of calibration scans were used to compute a set of system constants that describe the effects of system delays and eddy currents to enable accurate reconstruction of data collected before uncorrected eddy currents have decayed. The ability of the proposed technique to reproducibly characterize small fixed delays (<50 μs) and short‐time constant (<1 ms) eddy currents is demonstrated. Magn Reson Med, 2009.Keywords
This publication has 17 references indexed in Scilit:
- NMR probes for measuring magnetic fields and field dynamics in MR systemsMagnetic Resonance in Medicine, 2008
- Improved k‐space trajectory measurement with signal shiftingMagnetic Resonance in Medicine, 2007
- Centering the projection reconstruction trajectory: Reducing gradient delay errorsMagnetic Resonance in Medicine, 2003
- A Method to measure arbitrary k‐space trajectories for rapid MR imagingMagnetic Resonance in Medicine, 1997
- A novel eddy current compensation scheme for pulsed gradient systemsMagnetic Resonance in Medicine, 1994
- Correction of eddy-current-induced B0 shifts by receiver reference-phase modulationJournal of Magnetic Resonance (1969), 1992
- Selection of a convolution function for Fourier inversion using gridding (computerised tomography application)IEEE Transactions on Medical Imaging, 1991
- Analytical method for the compensation of eddy-current effects induced by pulsed magnetic field gradients in NMR systemsJournal of Magnetic Resonance (1969), 1990
- Optimization of eddy-current compensationJournal of Magnetic Resonance (1969), 1990
- Correction of distortions due to the pulsed magnetic field gradient‐induced shift in B0 field by postprocessingMagnetic Resonance in Medicine, 1989