Two instruments based on differential optical absorption spectroscopy (DOAS) to measure accurate ammonia concentrations in the atmosphere
Authors:
Open Access
- 21 February 2012
- journal article
- Published by Copernicus GmbH in Atmospheric Measurement Techniques
- Vol. 5 (2) , 413-427
Abstract
We present two Differential Optical Absorption Spectroscopy (DOAS) instruments built at RIVM: the RIVM DOAS and the miniDOAS. Both instruments provide virtually interference-free measurements of NH3 concentrations in the atmosphere, since they measure over an open path, without suffering from inlet problems or interference problems by ammonium aerosols dissociating on tubes or filters. They measure concentrations up to at least 200 μg m−3, have a fast response, low maintenance demands, and a high up-time. The RIVM DOAS has a high accuracy of typically 0.15 μg m−3 for ammonia for 5-min averages and over a total light path of 100 m. The miniDOAS has been developed for application in measurement networks such as the Dutch National Air Quality Monitoring Network (LML). Compared to the RIVM DOAS it has a similar accuracy, but is significantly reduced in size, costs, and handling complexity. The RIVM DOAS and miniDOAS results showed excellent agreement (R2 = 0.996) during a field measurement campaign in Vredepeel, the Netherlands. This measurement site is located in an agricultural area and is characterized by highly variable, but on average high ammonia concentrations in the air. The RIVM-DOAS and miniDOAS results were compared to the results of the AMOR instrument, a continuous-flow wet denuder system, which is currently used in the LML. Averaged over longer time spans of typically a day, the (mini)DOAS and AMOR results agree reasonably well, although an offset of the AMOR values compared to the (mini)DOAS results exists. On short time scales, the (mini)DOAS shows a faster response and does not show the memory effects due to inlet tubing and transport of absorption fluids encountered by the AMOR. Due to its high accuracy, high uptime, low maintenance and its open path, the (mini)DOAS shows a good potential for flux measurements by using two (or more) systems in a gradient set-up and applying the aerodynamic gradient technique.Keywords
All Related Versions
This publication has 18 references indexed in Scilit:
- Measurement of atmospheric ammonia at a dairy using differential optical absorption spectroscopy in the mid-ultravioletAtmospheric Environment, 2002
- Low and room temperature photoabsorption cross sections of NH3 in the UV regionPlanetary and Space Science, 1998
- Ammonia: emission, atmospheric transport and depositionNew Phytologist, 1998
- An automatic atmospheric ammonia network in the Netherlands set-up and resultsAtmospheric Environment, 1998
- The NO2 absorption spectrum. III: The 200?300 nm region at ambient temperatureJournal of Atmospheric Chemistry, 1996
- A compilation of the absorption cross-sections of SO2 from 106 to 403 nmJournal of Quantitative Spectroscopy and Radiative Transfer, 1993
- Differential optical absorption spectroscopy (DOAS) system for urban atmospheric pollution monitoringApplied Optics, 1993
- Absolute absorption cross sections of ozone in the 185‐ to 350‐nm wavelength rangeJournal of Geophysical Research: Atmospheres, 1986
- Absorption cross section measurements of oxygen in the wavelength region 195–241 nm of the Herzberg continuumPlanetary and Space Science, 1986
- Nitrous oxide ultraviolet absorption spectrum at stratospheric temperaturesGeophysical Research Letters, 1977