A Novel Slope Method for Measurement of Fluid Density with a Micro-cantilever under Flexural and Torsional Vibrations
Open Access
Abstract
A novel method, which was called a slope method, has been proposed to measure fluid density by the micro-cantilever sensing chip. The theoretical formulas of the slope method were discussed and established when the micro-cantilever sensing chip was under flexural and torsional vibrations. The slope was calculated based on the fitted curve between the excitation and output voltages of sensing chip under the nonresonant status. This measuring method need not sweep frequency to find the accurate resonant frequency. Therefore, the fluid density was measured easily based on the calculated slope. In addition, the micro-cantilver was drived by double sided excitation and free end excitation to oscillate under flexural and torsional vibrations, respectively. The corresponding experiments were carried out to measure the fluid density by the slope method. The measurement results were also analyzed when the sensing chip was under flexural and torsional nonresonant vibrations separately. The measurement accuracies under these vibrations were all better than 1.5%, and the density measuring sensitivity under torsional nonresonant vibration was about two times higher than that under flexural nonresonant vibration.Keywords
This publication has 21 references indexed in Scilit:
- Miniature density–viscosity measurement cell utilizing electrodynamic-acoustic resonator sensorsSensors and Actuators A: Physical, 2011
- Utilizing a high fundamental frequency quartz crystal resonator as a biosensor in a digital microfluidic platformSensors and Actuators A: Physical, 2011
- A MEMS Vibrating Edge Supported Plate for the Simultaneous Measurement of Density and Viscosity: Results for Argon, Nitrogen, and Methane at Temperatures from (297 to 373) K and Pressures between (1 and 62) MPaJournal of Chemical & Engineering Data, 2008
- Multi-parameter microcantilever sensor for comprehensive characterization of Newtonian fluidsSensors and Actuators B: Chemical, 2008
- Density of Diethyl Adipate using a New Vibrating Tube Densimeter from (293.15 to 403.15) K and up to 140 MPa. Calibration and MeasurementsJournal of Chemical & Engineering Data, 2008
- A silicon straight tube fluid density sensorJournal of Micromechanics and Microengineering, 2007
- Viscosity and density values from excitation level response of piezoelectric-excited cantilever sensorsSensors and Actuators A: Physical, 2007
- Viscosity-induced errors in the density determination of room temperature ionic liquids using vibrating tube densitometryFluid Phase Equilibria, 2007
- Sensitivity of flexural and torsional vibration modes of atomic force microscope cantilevers to surface stiffness variationsNanotechnology, 2001
- A low-pressure encapsulated resonant fluid density sensor with feedback control electronicsMeasurement Science and Technology, 2000