Effects of time delay and blood storage methods on analysis of canine venous blood samples with an Element point‐of‐care analyzer
- 27 December 2022
- journal article
- research article
- Published by Wiley in Veterinary Clinical Pathology
- Vol. 52 (1) , 22-29
Abstract
BackgroundManufacturers of point-of-care (POC) analyzers recommend immediate processing and anaerobic collection of blood samples. However, it is not uncommon for clinical scenarios to result in delayed sample processing or room air exposure that could impact the test results. ObjectiveTo investigate the effect of time delay and sample storage method on key POC analytes in canine venous blood samples processed with an Element POC analyzer. MethodsBlood gas analysis was performed on venous blood samples at times 0 (T0), 15, 30, and 60 minutes after sampling using three different storage methods: preheparinized plastic syringes and two different lithium heparin tubes. To determine clinical relevance, results were compared with allowable total error of the respective parameter. Significance was set at P < 0.05. ResultsSignificant differences between the three storage methods at baseline were found for partial pressure of carbon dioxide (PCO2), partial pressure of oxygen (PO2), base excess, and total hemoglobin. No significant differences up to T60 were found within collection methods for actual bicarbonate (HCO3-), base excess, sodium, potassium, chloride, ionized calcium (iCa), glucose, and BUN. Significant differences within collection methods were found after T0 for creatinine, after 15 minutes for lactate, and after 30 minutes for pH and hematocrit. No significant differences were found for PO2 in samples stored in preheparinized plastic syringes at any time point. ConclusionsThese results suggest that HCO3-, sodium, potassium, chloride, iCa, glucose, and BUN are comparable within the three storage methods for up to 60 minutes after sampling without resulting in clinically relevant changes.Keywords
This publication has 17 references indexed in Scilit:
- Blood Gas AnalyzersTopics in Companion Animal Medicine, 2016
- Effect of Sample Storage Temperature and Time Delay on Blood Gases, Bicarbonate and pH in Human Arterial Blood SamplesIranian Red Crescent Medical Journal, 2015
- Stability of 35 biochemical and immunological routine tests after 10 hours storage and transport of human whole blood at 21°CScandinavian Journal of Clinical and Laboratory Investigation, 2014
- Comparison of theEPOCand i‐STATanalysers for canine blood gas and electrolyte analysisJournal of Small Animal Practice, 2014
- ASVCPguidelines: allowable total error guidelines for biochemistryVeterinary Clinical Pathology, 2013
- ASVCP guidelines: quality assurance for point‐of‐care testing in veterinary medicineVeterinary Clinical Pathology, 2013
- Effect of sample handling on venous PCO2, pH, bicarbonate, and base excess measured with a point‐of‐care analyzerJournal of Veterinary Emergency and Critical Care, 2004
- Current databases on biological variation: pros, cons and progressScandinavian Journal of Clinical and Laboratory Investigation, 1999
- ABC of oxygen: Assessing and interpreting arterial blood gases and acid-base balanceBMJ, 1998
- Stability of pO 2 , pCO 2 and pH in Heparinized Whole Blood Samples: Influence of Storage Temperature with Regard to Leukocyte Count and Syringe MaterialClinical Chemistry and Laboratory Medicine (CCLM), 1992